:ANALysis:PANalyze<x>:SWLoss:ILEVel
Function Sets or queries the current level used to determine
the zero loss period for total loss computation.
Syntax
:ANALysis:PANalyze<x>:SWLoss:ILEV
el {<NRf>|<Current>}
:ANALysis:PANalyze<x>:SWLoss:ILEVel?
<x> = 1 or 2
<NRf>, <Current> = See the DLM2000 Features
Guide for this information.
Example
:ANALYSIS:PANALYZE1:SWLOSS:ILEVEL 1
:ANALYSIS:PANALYZE1:SWLOSS:ILEVEL?
-> :ANALYSIS:PANALYZE1:SWLOSS:
ILEVEL 1.0000000E+00
:ANALysis:PANalyze<x>:SWLoss:MATH
Function Sets or queries the on/off status of the power
waveform display.
Syntax
:ANALysis:PANalyze<x>:SWLoss:MATH {OF
F|POWer}
:ANALysis:PANalyze<x>:SWLoss:MATH?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:MATH OFF
:ANALYSIS:PANALYZE1:SWLOSS:MATH?
-> :ANALYSIS:PANALYZE1:SWLOSS:MATH OF
F
:ANALysis:PANalyze<x>:SWLoss:MEASu
re?
Function Queries all of the settings for the automated
measurement of power supply analysis
parameters.
Syntax
:ANALysis:PANalyze<x>:SWLoss:MEASure?
<x> = 1 or 2
:ANALysis:PANalyze<x>:SWLoss:MEASure
:<Parameter>?
Function Queries the setting of a power supply analysis
parameter.
Syntax
:ANALysis:PANalyze<x>:SWLoss:MEASure:
<Parameter>?
<x> = 1 or 2
<Parameter> = {P|PABS|PN|PP|WH|WHABs|
WHN|WHP|Z}
:ANALysis:PANalyze<x>:SWLoss:DPROxim
al:PERCent
Function Sets or queries the distal, mesial, and proximal
points as percentages.
Syntax
:ANALysis:PANalyze<x>:SWLoss:DPROxima
l:PERCent {<NRf>,<NRf>,<NRf>}
:ANALysis:PANalyze<x>:SWLoss:DPROxima
l:PERCent?
<NRf>, <NRf>, <NRf> = 0 to 100 (%, in steps of 1)
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:DPROXIMAL:
PERCENT 10,50,90
:ANALYSIS:PANALYZE1:SWLOSS:DPROXIMAL:
PERCENT?
-> :ANALYSIS:PANALYZE1:SWLOSS:DPROXIM
AL:PERCENT 10,50,90
:ANALysis:PANalyze<x>:SWLoss:DPROxim
al:UNIT
Function Sets or queries the distal, mesial, and proximal
points as voltages.
Syntax
:ANALysis:PANalyze<x>:SWLoss:DPROxima
l:UNIT {<NRf>,<NRf>,<NRf>}
:ANALysis:PANalyze<x>:SWLoss:DPROxima
l:UNIT?
<NRf>, <NRf>, <NRf> = See the DLM2000
Features Guide for this information.
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:DPROXIMAL:
UNIT−1,0,1
:ANALYSIS:PANALYZE1:SWLOSS:DPROXIMAL:
UNIT?
-> :ANALYSIS:PANALYZE1:SWLOSS:DPROXIM
AL:UNIT−1.0000000E+00,
0.0000000E+00,1.0000000E+00
:ANALysis:PANalyze<x>:SWLoss:DTYPe
Function Sets or queries the device type for total loss
computation.
Syntax
:ANALysis:PANalyze<x>:SWLoss:DTY
Pe {IGBT|MOSFET|OFF}
:ANALysis:PANalyze<x>:SWLoss:DTYPe?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:DTYPE IGBT
:ANALYSIS:PANALYZE1:SWLOSS:DTYPE?
-> :ANALYSIS:PANALYZE1:SWLOSS:DTYPE