5-72
IM DLM4038-17EN
:ANALysis:PANalyze<x>:SWLoss:MEASure
:<Parameter>:VALue?
Function Queries an automatically measured value of a
power supply analysis parameter.
Syntax
:ANALysis:PANalyze<x>:SWLoss:
MEASure:<Parameter>:VALue? [{<NRf>}]
<x> = 1 to 2
<Parameter> =
{P|PABS|PN|PP|WH|WHABs|
WHN|WHP|Z}
<NRf> = 1 to 100000
Example
:ANALYSIS:PANALYZE1:SWLOSS:MEASURE:
P:VALUE?
-> :ANALYSIS:PANALYZE1:SWLOSS:MEASUR
E:P:VALUE 10.0000E+03
Description If the value is immeasurable, the DLM4000
returns “NAN” (not a number).
The <NRf> is used to specify which iteration of
automated measurement to query the measured
value from.
If <NRf> is set to 1, the oldest measured value
in the automated measurement memory is
queried.
If a measured value does not exist at the
specified iteration, the DLM4000 returns “NAN”
(not a number).
If <NRf> is omitted, the most recent measured
value is queried.
If cyclic statistical processing for automated
measurement is being executed and <NRf> is
specified, the measured values over a cycle in
iteration <NRf> from the left of the display on
the displayed waveform is queried. If <NRf>
is omitted, the measured values over the last
cycle on the displayed waveform is queried.
:ANALysis:PANalyze<x>:SWLoss:METhod
Function Sets or queries the calculation method for high
and low points.
Syntax
:ANALysis:PANalyze<x>:SWLoss:METh
od {AUTO|MAXimum|HISTogram}
:ANALysis:PANalyze<x>:SWLoss:METhod?
Example
:ANALYSIS:PANALYZE1:SWLOSS:METHOD AU
TO
:ANALYSIS:PANALYZE1:SWLOSS:METHOD?
-> :ANALYSIS:PANALYZE1:SWLOSS:METHO
D AUTO
:ANALysis:PANalyze<x>:SWLoss:RANGe
Function Sets or queries the measurement source window.
Syntax
:ANALysis:PANalyze<x>:SWLoss:RAN
Ge {MAIN|Z1|Z2}
:ANALysis:PANalyze<x>:SWLoss:RANGe?
<x> = 1 to 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:RAN
GE MAIN
:ANALYSIS:PANALYZE1:SWLOSS:RANGE?
-> :ANALYSIS:PANALYZE1:SWLOSS:RAN
GE MAIN
:ANALysis:PANalyze<x>:SWLoss:MEASure
:<Parameter>:{MAXimum|MEAN|MINimum|S
DEViation}?
Function Queries a statistical value of a power supply
analysis parameter.
Syntax
:ANALysis:PANalyze<x>:SWLoss:
MEASure:<Parameter>:{MAXimum|MEAN|
MINimum|SDEViation}?
<x> = 1 to 2
<Parameter> =
{P|PABS|PN|PP|WH|WHABs|
WHN|WHP|Z}
Example
:ANALYSIS:PANALYZE1:SWLOSS:MEASURE:
P:MAXIMUM?
-> :ANALYSIS:PANALYZE1:SWLOSS:MEASUR
E:P:MAXIMUM 1.000E+00
Description If the statistical value is immeasurable, the
DLM4000 returns “NAN” (not a number).
:ANALysis:PANalyze<x>:SWLoss:MEASure
:<Parameter>:STATe
Function Sets or queries the on/off status of a power
supply analysis parameter.
Syntax
:ANALysis:PANalyze<x>:SWLoss:MEASure
:<Parameter>:STATe {<Boolean>}
:ANALysis:PANalyze<x>:SWLoss:MEASure
:<Parameter>:STATe?
<x> = 1 to 2
<Parameter> =
{P|PABS|PN|PP|WH|WHABs|
WHN|WHP|Z}
Example
:ANALYSIS:PANALYZE1:SWLOSS:MEASURE:
P:STATE ON
:ANALYSIS:PANALYZE1:SWLOSS:MEASURE:
P:STATE?
-> :ANALYSIS:PANALYZE1:SWLOSS:MEASUR
E:P:STATE 1
5.3 ANALysis Group

Table of Contents

Other manuals for YOKOGAWA DLM4038

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM4038 and is the answer not in the manual?

YOKOGAWA DLM4038 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM4038
CategoryTest Equipment
LanguageEnglish

Related product manuals