:ANALysis:PANalyze<x>:SWLoss:RDS
Function Sets or queries the on-resistance value for total
loss computation.
Syntax
:ANALysis:PANalyze<x>:SWLoss:R
DS {<NRf>}
:ANALysis:PANalyze<x>:SWLoss:RDS?
<x> = 1 to 2
<NRf> = 0 to 100 (in 1 mΩ steps)
Example
:ANALYSIS:PANALYZE1:SWLOSS:RDS 1
:ANALYSIS:PANALYZE1:SWLOSS:RDS?
-> :ANALYSIS:PANALYZE1:SWLOSS:RDS 1.
000E+00
:ANALysis:PANalyze<x>:SWLoss:SCALe?
Function Queries all scaling settings.
Syntax
:ANALysis:PANalyze<x>:SWLoss:SCALe?
<x> = 1 to 2
:ANALysis:PANalyze<x>:SWLoss:SCALe:C
ENTer
Function Sets or queries the level of the center position for
manual scaling.
Syntax
:ANALysis:PANalyze<x>:SWLoss:SCALe:C
ENTer {<NRf>}
:ANALysis:PANalyze<x>:SWLoss:SCALe:C
ENTer?
<x> = 1 to 2
<NRf> = −1.0000E+31 to 1.0000E+31
Example
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:CEN
TER 1
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:CEN
TER?
-> :ANALYSIS:PANALYZE1:SWLOSS:SCALE:
CENTER 1.00000E+00
:ANALysis:PANalyze<x>:SWLoss:SCALe:M
ODE
Function Sets or queries the scaling mode.
Syntax
:ANALysis:PANalyze<x>:SWLoss:SCALe:M
ODE {AUTO|MANual}
:ANALysis:PANalyze<x>:SWLoss:SCALe:M
ODE?
<x> = 1 to 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:MO
DE AUTO
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:MO
DE?
-> :ANALYSIS:PANALYZE1:SWLOSS:SCALE:
MODE AUTO
:ANALysis:PANalyze<x>:SWLoss:SCALe:S
ENSitivity
Function Sets or queries the sensitivity of the center
position for manual scaling.
Syntax
:ANALysis:PANalyze<x>:SWLoss:SCALe:S
ENSitivity {<NRf>}
:ANALysis:PANalyze<x>:SWLoss:SCALe:S
ENSitivity?
<x> = 1 to 2
<NRf> = −1.0000E+31 to 1.0000E+31
Example
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:SEN
SITIVITY 10
:ANALYSIS:PANALYZE1:SWLOSS:SCALE:SEN
SITIVITY?
-> :ANALYSIS:PANALYZE1:SWLOSS:SCALE:
SENSITIVITY 10.0000E+00
:ANALysis:PANalyze<x>:SWLoss:TRANge
(Time Range)
Function Sets or queries the measurement time period.
Syntax
:ANALysis:PANalyze<x>:SWLoss:TRAN
ge {<NRf>,<NRf>}
:ANALysis:PANalyze<x>:SWLoss:TRANge?
<x> = 1 to 2
<NRf> = −5 to 5 divisions (in steps of 10
divisions/display record length)
Example
:ANALYSIS:PANALYZE1:SWLOSS:
TRANGE4,−4
:ANALYSIS:PANALYZE1:SWLOSS:TRANGE?
-> :ANALYSIS:PANALYZE1:SWLOSS:TRANG
E4.00E+00,−4.00E+00
:ANALysis:PANalyze<x>:SWLoss:ULEVel
Function Sets or queries the voltage level used to
determine the loss calculation period for total loss
computation.
Syntax
:ANALysis:PANalyze<x>:SWLoss:ULEV
el {<NRf>|<Voltage>}
:ANALysis:PANalyze<x>:SWLoss:ULEVel?
<x> = 1 to 2
<NRf>, <Voltage> = See the DLM4000 Features
Guide for this information.
Example
:ANALYSIS:PANALYZE1:SWLOSS:ULEVEL 1V
:ANALYSIS:PANALYZE1:SWLOSS:ULEVEL?
-> :ANALYSIS:PANALYZE1:SWLOSS:ULEVE
L 1.0000000E+00
:ANALysis:PANalyze<x>:SWLoss:UNIT
Function Sets or queries the power unit.
Syntax
:ANALysis:PANalyze<x>:SWLoss:UNIT {J
|WH}
:ANALysis:PANalyze<x>:SWLoss:UNIT?
<x> = 1 to 2
Example
:ANALYSIS:PANALYZE1:SWLOSS:UNIT WH
:ANALYSIS:PANALYZE1:SWLOSS:UNIT?
-> :ANALYSIS:PANALYZE1:SWLOSS:UNIT W
H
5.3 ANALysis Group

Table of Contents

Other manuals for YOKOGAWA DLM4038

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM4038 and is the answer not in the manual?

YOKOGAWA DLM4038 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM4038
CategoryTest Equipment
LanguageEnglish

Related product manuals