Page #142 background imageLoading...
Page #142 background image
5-109
IM DLM4038-17EN
Commands
1
2
3
4
5
6
App
Index
:GONogo:NGStopcount
Function Sets or queries the number of NO-GO results
at which the DLM4000 will stop performing
determinations.
Syntax
:GONogo:NGStopcount {<NRf>|INFinity}
:GONogo:NGStopcount?
<NRf> = 1 to 1000
Example
:GONOGO:NGSTOPCOUNT 100
:GONOGO:NGSTOPCOUNT?
-> :GONOGO:NGSTOPCOUNT 100
:GONogo:STOPcount
Function Sets or queries the number of acquisitions
at which the DLM4000 will stop performing
determinations.
Syntax
:GONogo:STOPcount {<NRf>|INFinity}
:GONogo:STOPcount?
<NRf> = 1 to 1000000
Example
:GONOGO:STOPCOUNT 1000
:GONOGO:STOPCOUNT?
-> :GONOGO:STOPCOUNT 1000
:GONogo:WAIT?
Function Waits for the completion of GO/NO-GO
determination by using a timeout timer.
Syntax
:GONogo:WAIT? {<NRf>}
<NRf> = 0 to 864000 (in 100-ms steps)
Example
(Set the timeout to 5 seconds)
:GONOGO:WAIT? 50
-> :GONOGO:WAIT 0
Description The DLM4000 returns 0 if the operation finishes
before the timer expires and returns 1 if the timer
expires.
:GONogo:ZPARameter?
Function Queries all zone and parameter determination
settings.
Syntax
:GONogo:ZPARameter?
:GONogo[:ZPARameter]:NUMBer<x>?
Function Queries all of the settings related to the specified
reference standard for zone or parameter
determination.
Syntax
:GONogo[:ZPARameter]:NUMBer<x>?
<x> = 1 to 4
:GONogo[:ZPARameter]:NUMBer<x>:CAU
Se?
Function Queries whether or not the specified waveform
parameter is the cause of a NO-GO zone or
parameter determination.
Syntax
:GONogo[:ZPARameter]:NUMBer<x>:CAU
Se?
<x> = 1 to 4
Example
:GONOGO:ZPARAMETER:NUMBER1:CAUSE?
-> :GONOGO:ZPARAMETER:NUMBER1:CAU
SE 1
Description When the parameter is the cause of a NO-GO
result, the DLM4000 returns 1. Otherwise, the
DLM4000 returns 0.
:GONogo[:ZPARameter]:NUMBer<x>:CONDi
tion
Function Sets or queries the specified waveform
parameter’s reference condition for zone or
parameter determination.
Syntax
:GONogo[:ZPARameter]:NUMBer<x>:CONDi
tion {IN|OFF|OUT}
:GONogo[:ZPARameter]:NUMBer<x>:CONDi
tion?
<x> = 1 to 4
Example
:GONOGO:ZPARAMETER:NUMBER1:CONDITI
ON IN
:GONOGO:ZPARAMETER:NUMBER1:CONDITI
ON?
-> :GONOGO:ZPARAMETER:NUMBER1:CONDIT
ION IN
:GONogo[:ZPARameter]:NUMBer<x>:MODE
Function Sets or queries the specified reference standard
mode.
Syntax
:GONogo[:ZPARameter]:NUMBer<x>:
MODE {PARameter|POLYgon|RECTangle|
WAVE}
:GONogo[:ZPARameter]:NUMBer<x>:MODE?
<x> = 1 to 4
Example
:GONOGO:ZPARAMETER:NUMBER1:MODE PARA
METER
:GONOGO:ZPARAMETER:NUMBER1:MODE?
-> :GONOGO:ZPARAMETER:NUMBER1:MODE P
ARAMETER
:GONogo[:ZPARameter]:NUMBer<x>:PARam
eter?
Function Queries all parameter settings for the specified
reference standard.
Syntax
:GONogo[:ZPARameter]:NUMBer<x>:PARam
eter?
<x> = 1 to 4
5.14 GONogo Group

Table of Contents

Other manuals for YOKOGAWA DLM4038