5-110
IM DLM4038-17EN
:GONogo[:ZPARameter]:NUMBer<x>:PARam
eter:ITEM
Function Sets or queries the specified waveform
parameter’s item for parameter determination.
Syntax
:GONogo[:ZPARameter]:NUMBer<x>:PARam
eter:ITEM {<Parameter>}
:GONogo[:ZPARameter]:NUMBer<x>:PARam
eter:ITEM?
<x> = 1 to 4
<Parameter> =
{AMPLitude|AVERage|
AVGFreq|AVGPeriod|BWIDth|DELay|DT|
DUTYcycle|ENUMber|FALL|FREQuency|
HIGH|LOW|MAXimum|MINimum|
NOVershoot|NWIDth|PERiod|PNUMber|
POVershoot|PTOPeak|PWIDth|RISE|RMS|
SDEViation|TY1Integ|TY2Integ|V1|V2}
Example
:GONOGO:ZPARAMETER:NUMBER1:
PARAMETER:ITEM AVERAGE
:GONOGO:ZPARAMETER:NUMBER1:
PARAMETER:ITEM?
-> :GONOGO:ZPARAMETER:NUMBER1:PARAME
TER:ITEM AVERAGE
:GONogo[:ZPARameter]:NUMBer<x>:PARam
eter:LIMit
Function Sets or queries the specified waveform
parameter’s upper and lower limits for parameter
determination.
Syntax
:GONogo[:ZPARameter]:NUMBer<x>:PARam
eter:LIMit {<NRf>}
:GONogo[:ZPARameter]:NUMBer<x>:PARam
eter:LIMit?
<x> = 1 to 4
<NRf> = See the DLM4000 Features Guide for
this information.
Example
:GONOGO:ZPARAMETER:NUMBER1:
PARAMETER:LIMIT 0,1
:GONOGO:ZPARAMETER:NUMBER1:
PARAMETER:LIMIT?
-> :GONOGO:ZPARAMETER:NUMBER1:PARAME
TER:LIMIT 1.000E+00,0.000E+00
:GONogo[:ZPARameter]:NUMBer<x>:PARam
eter:TRACe
Function Sets or queries the specified waveform
parameter’s source waveform for parameter
determination.
Syntax
:GONogo[:ZPARameter]:NUMBer<x>:PARam
eter:TRACe {<NRf>|FFT<y>|MATH<y>|
PODA<y>|PODB<y>|PODL<y>|XY<y>}
:GONogo[:ZPARameter]:NUMBer<x>:PARam
eter:TRACe?
<x> = 1 to 4
<NRf> = 1 to 8
<y> of FFT<y> = 1 to 2
<y> of MATH<y> = 1 to 4
<y> of PODA<y>, PODB<y>, PODL<y> = 0 to 7
<y> of XY<y> = 1 to 4
Example
:GONOGO:ZPARAMETER:NUMBER1:
PARAMETER:TRACE 1
:GONOGO:ZPARAMETER:NUMBER1:
PARAMETER:TRACE?
-> :GONOGO:ZPARAMETER:NUMBER1:PARAME
TER:TRACE 1
:GONogo[:ZPARameter]:NUMBer<x>:PARam
eter:VALue?
Function Queries the specified waveform parameter’s
measured value for parameter determination.
Syntax
:GONogo[:ZPARameter]:NUMBer<x>:PARam
eter:VALue?
<x> = 1 to 4
Example
:GONOGO:ZPARAMETER:NUMBER1:
PARAMETER:VALUE?
-> :GONOGO:ZPARAMETER:NUMBER1:PARAME
TER:VALUE1.98E−03
Description If
:GONogo[:ZPARameter]:NUMBer<x>:C
ONDition
is set to OFF or when the value is
otherwise immeasurable, the DLM4000 returns
“NAN” (not a number).
:GONogo[:ZPARameter]:NUMBer<x>:POLYg
on?
Function Queries all polygonal zone determination settings.
Syntax
:GONogo[:ZPARameter]:NUMBer<x>:POLYg
on?
<x> = 1 to 4
5.14 GONogo Group

Table of Contents

Other manuals for YOKOGAWA DLM4038