Commands
1
2
3
4
5
6
App
Index
:MEASure:USER<x>:STATe
Function Sets or queries the on/off status of automated
measurement of a Calc item.
Syntax
:MEASure:USER<x>:STATe {<Boolean>}
:MEASure:USER<x>:STATe?
<x> = 1 to 4
Example
:MEASURE:USER1:STATE ON
:MEASURE:USER1:STATE?
-> :MEASURE:USER1:STATE 1
:MEASure:USER<x>:UNIT
Function Sets or queries the unit of a Calc item.
Syntax
:MEASure:USER:UNIT {<String>}
:MEASure:USER:UNIT?
<String> = Up to 4 characters
Example
:MEASURE:USER1:UNIT "ABC"
:MEASURE:USER1:UNIT?
-> :MEASURE:USER1:UNIT "ABC"
:MEASure:USER<x>:VALue?
Function Queries the automatically measured value of a
Calc item.
Syntax
:MEASure:USER<x>:VALue? [<NRf>]
<x> = 1 to 4
<NRf> = 1 to 100000
Example
:MEASURE:USER1:VALUE?
-> :MEASURE:USER1:VALUE 1.000E+00
Description If the value is immeasurable, the DLM4000
returns “NAN” (not a number).
The <NRf> is used to specify which iteration of
automated measurement to query the measured
value from.
If <NRf> is set to 1, the oldest measured value
in the automated measurement memory is
queried.
If a measured value does not exist at the
specified iteration, the DLM4000 returns “NAN”
(not a number).
If <NRf> is omitted, the most recent measured
value is queried.
:MEASure:WAIT?
Function Waits for the completion of waveform parameter
automated measurement with a set timeout.
Syntax
:MEASure:WAIT? {<NRf>}
<NRf> = 1 to 36000 (the timeout specified in
100 ms intervals)
Example
:MEASURE:WAIT? 1
-> :MEASURE:WAIT 0
Description The command returns zero if the automated
measurement finishes within the specified
timeout. If automated measurement does not
finish, or if it was never taking place to begin
with, the command returns 1.
Even if you set a long timeout, the command will
return zero as soon as automated measurement
finishes.
5.21 MEASure Group

Table of Contents

Other manuals for YOKOGAWA DLM4038

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM4038 and is the answer not in the manual?

YOKOGAWA DLM4038 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM4038
CategoryTest Equipment
LanguageEnglish

Related product manuals