5-154
IM DLM4038-17EN
:SEARch:SKIP:HOLDoff
Function Sets or queries how long searching will be
skipped.
Syntax
:SEARch:SKIP:HOLDoff {<Time>}
:SEARch:SKIP:HOLDoff?
<Time> = 0.1 ns to 1 s
Example
:SEARCH:SKIP:HOLDOFF 0S
:SEARCH:SKIP:HOLDOFF?
-> :SEARCH:SKIP:HOLDOFF 0.000E+00
:SEARch:SKIP:MODE
Function Sets or queries the skip mode.
Syntax
:SEARch:SKIP:MODE {DECimation|
HOLDoff|OFF}
:SEARch:SKIP:MODE?
Example
:SEARCH:SKIP:MODE DECIMATION
:SEARCH:SKIP:MODE?
-> :SEARCH:SKIP:MODE DECIMATION
Description This command is valid only when
:SEARch:TYPE
is set to EDGE or PWIDth.
:SEARch:SPOint
Function Sets or queries search start point.
Syntax
:SEARch:SPOint {<NRf>}
:SEARch:SPOint?
<NRf> = −5 to 5 divisions (in steps of 10
divisions/display record length)
Example
:SEARCH:SPOINT−1
:SEARCH:SPOINT?
->:SEARCH:SPOINT−1.000E+00
:SEARch:TWINdow
Function Sets or queries which window will display the
detected points.
Syntax
:SEARch:TWINdow {Z1|Z2}
:SEARch:TWINdow?
Example
:SEARCH:TWINDOW Z1
:SEARCH:TWINDOW?
-> :SEARCH:TWINDOW Z1
:SEARch:TYPE
Function Sets or queries search type.
Syntax
:SEARch:TYPE {EDGE|QUALify|PWIDth|
STATe|SWIDth}
:SEARch:TYPE?
Example
:SEARCH:TYPE EDGE
:SEARCH:TYPE?
-> :SEARCH:TYPE EDGE
:SEARch:WIDTh?
Function Queries all pulse width search settings.
Syntax
:SEARch:WIDTh?
:SEARch:WIDTh:HYSTeresis
Function Sets or queries the pulse width search level
hysteresis.
Syntax
:SEARch:WIDTh:HYSTeresis {<NRf>}
:SEARch:WIDTh:HYSTeresis?
<NRf> = 0.3 to 4 divisions in 0.1 steps
Example
:SEARCH:WIDTH:HYSTERESIS 1
:SEARCH:WIDTH:HYSTERESIS?
-> :SEARCH:WIDTH:HYSTERESIS 1.0
:SEARch:WIDTh:LEVel
Function Sets or queries the pulse width search level.
Syntax
:SEARch:WIDTh:LEVel {<NRf>|
<Voltage>}
:SEARch:WIDTh:LEVel?
<NRf> = See the DLM4000 Features Guide for
this information.
Example
:SEARCH:WIDTH:LEVEL 1V
:SEARCH:WIDTH:LEVEL?
-> :SEARCH:WIDTH:LEVEL 1.000E+00
Description When the probe type is set to current, this
command sets or queries the <Current> value.
:SEARch:WIDTh:POLarity
Function Sets or queries the pulse width search level
polarity.
Syntax
:SEARch:WIDTh:POLarity {HIGH|LOW}
:SEARch:WIDTh:POLarity?
Example
:SEARCH:WIDTH:POLARITY HIGH
:SEARCH:WIDTH:POLARITY?
-> :SEARCH:WIDTH:POLARITY HIGH
:SEARch:WIDTh:SOURce
Function Sets or queries the pulse width search source
waveform.
Syntax
:SEARch:WIDTh:SOURce {<NRf>|MATH<x>|
NONE|PODA<x>|PODB<x>|PODL<x>}
:SEARch:WIDTh:SOURce?
<NRf> = 1 to 8
<x> of MATH<x> = 1 to 4
<x> of PODA<x>, PODB<x>, PODL<x> = 0 to 7
Example
:SEARCH:WIDTH:SOURCE 1
:SEARCH:WIDTH:SOURCE?
-> :SEARCH:WIDTH:SOURCE 1
:SEARch:WIDTh:TIME<x>
Function Sets or queries the pulse width for pulse width
searching.
Syntax
:SEARch:WIDTh:TIME<x> {<Time>}
:SEARch:WIDTh:TIME<x>?
<x> = 1 to 2
<Time> = 1 ns to 500 s
Example
:SEARCH:WIDTH:TIME1 1S
:SEARCH:WIDTH:TIME1?
-> :SEARCH:WIDTH:TIME1 1.000E+00
5.24 SEARch Group

Table of Contents

Other manuals for YOKOGAWA DLM4038

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM4038 and is the answer not in the manual?

YOKOGAWA DLM4038 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM4038
CategoryTest Equipment
LanguageEnglish

Related product manuals