5-168
IM DLM4038-17EN
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:ACK:MODE
Function Sets or queries the ACK mode, which is one
of the ID and data conditions, for CAN FD bus
signal searching.
Syntax
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:ACK:MODE {<Boolean>}
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:ACK:MODE?
<x> = 1 to 4
Example
:SERIALBUS1:CANFD:SEARCH:SETUP:IDDAT
A:ACK:MODE ON
:SERIALBUS1:CANFD:SEARCH:SETUP:IDDAT
A:ACK:MODE?
-> :SERIALBUS1:CANFD:SEARCH:SETUP:ID
DATA:ACK:MODE 1
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:ACK:TYPE
Function Sets or queries the ACK condition, which is one
of the ID and data conditions, for CAN FD bus
signal searching.
Syntax
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:ACK:TYPE {ACK|ACKBoth|NONack}
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:ACK:TYPE?
<x> = 1 to 4
Example
:SERIALBUS1:CANFD:SEARCH:SETUP:IDDAT
A:ACK:TYPE ACK
:SERIALBUS1:CANFD:SEARCH:SETUP:IDDAT
A:ACK:TYPE?
-> :SERIALBUS1:CANFD:SEARCH:SETUP:ID
DATA:ACK:TYPE ACK
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:DATA?
Function Queries all data settings for the ID and data
conditions for CAN FD bus signal searching.
Syntax
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:DATA?
<x> = 1 to 4
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:DATA:BCOunt
Function Sets or queries the pattern comparison position,
which is one of the ID and data conditions, for
CAN FD bus signal searching.
Syntax
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:DATA:BCOUNT {<NRf>}
<x> = 1 to 4
<NRf> = 0 to 63
Example
:SERIALBUS1:CANFD:SEARCH:SETUP:IDDAT
A:DATA:BCOUNT 4
:SERIALBUS1:CANFD:SEARCH:SETUP:IDDAT
A:DATA:BCOUNT?
-> :SERIALBUS1:CANFD:SEARCH:SETUP:ID
DATA:DATA:BCOUNT 4
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:DATA:CONDition
Function Sets or queries the comparison condition, which
is one of the ID and data conditions, for CAN FD
bus signal searching.
Syntax
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:DATA:CONDition {BETWeen|EQUal|
FALSe|GREater|LESS|NOTBetween|NOTEqu
l|TRUE}
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:DATA:CONDition?
<x> = 1 to 4
Example
:SERIALBUS1:CANFD:SEARCH:SETUP:IDDAT
A:DATA:CONDITION BETWEEN
:SERIALBUS1:CANFD:SEARCH:SETUP:IDDAT
A:DATA:CONDITION?
-> :SERIALBUS1:CANFD:SEARCH:SETUP:ID
DATA:DATA:CONDITION BETWEEN
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:DATA:DBYTe
Function Sets or queries the number of data bytes, which
is one of the ID and data conditions, for CAN FD
bus signal searching.
Syntax
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:DATA:DBYTe {<NRf>}
<x> = 1 to 4
<NRf> = 0 to 8
Example
:SERIALBUS1:CANFD:SEARCH:SETUP:IDDAT
A:DATA:DBYTE 2
:SERIALBUS1:CANFD:SEARCH:SETUP:IDDAT
A:DATA:DBYTE?
-> :SERIALBUS1:CANFD:SEARCH:SETUP:ID
DATA:DATA:DBYTE 2
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:DATA:DECimal<y>
Function Sets a reference value, which is one of the ID and
data conditions, for CAN FD bus signal searching
in decimal notation.
Syntax
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:DATA:DECimal<y> {<NRf>}
:SERialbus<x>:CANFD:SEARch:SETup[:ID
Data]:DATA:DECimal<y>?
<x> = 1 to 4
<y> = 1 or 2
<NRf> = –9E+18 to 9E+18
Example
:SERIALBUS1:CANFD:SEARCH:SETUP:IDDAT
A:DATA:DECIMAL1 1000
:SERIALBUS1:CANFD:SEARCH:SETUP:IDDAT
A:DATA:DECIMAL1?
-> :SERIALBUS1:CANFD:SEARCH:SETUP:ID
DATA:DATA:DECIMAL1 1.0000000E+03