EasyManua.ls Logo

YOKOGAWA DLM4038 - Page 227

YOKOGAWA DLM4038
373 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-194
IM DLM4038-17EN
:SERialbus<x>:I2C:SEARch[:SETup]:MO
DE
Function Sets or queries the I
2
C bus signal search type.
Syntax
:SERialbus<x>:I2C:SEARch[:SETup]:MO
DE {ADRData|EVERystart|GENeralcall|H
SMode|NONack|STARtbyte}
:SERialbus<x>:I2C:SEARch[:SETup]:MO
DE?
<x> = 1 to 4
Example
:SERIALBUS1:I2C:SEARCH:SETUP:MODE AD
RDATA
:SERIALBUS1:I2C:SEARCH:SETUP:MODE?
-> :SERIALBUS1:I2C:SEARCH:SETUP:
MODE ADRDATA
:SERialbus<x>:I2C:SEARch[:SETup]:NON
ack?
Function Queries all NON ACK ignore mode settings for
I
2
C bus signal searching.
Syntax
:SERialbus<x>:I2C:SEARch[:SETup]:NON
ack?
<x> = 1 to 4
:SERialbus<x>:I2C:SEARch[:SETup]:NON
ack:HSMode
Function Sets or queries whether or not NON ACK will be
ignored in high-speed mode for I
2
C bus signal
searching.
Syntax
:SERialbus<x>:I2C:SEARch[:SETup]:NON
ack:HSMode {<Boolean>}
:SERialbus<x>:I2C:SEARch[:SETup]:NON
ack:HSMode?
<x> = 1 to 4
Example
:SERIALBUS1:I2C:SEARCH:SETUP:NONACK:
HSMODE ON
:SERIALBUS1:I2C:SEARCH:SETUP:NONACK:
HSMODE?
-> :SERIALBUS1:I2C:SEARCH:SETUP:NONA
CK:HSMODE 1
:SERialbus<x>:I2C:SEARch[:SETup]:NON
ack:READaccess
Function Sets or queries whether or not NON ACK will be
ignored in read access mode for I
2
C bus signal
searching.
Syntax
:SERialbus<x>:I2C:SEARch[:SETup]:NON
ack:READaccess {<Boolean>}
:SERialbus<x>:I2C:SEARch[:SETup]:NON
ack:READaccess?
<x> = 1 to 4
Example
:SERIALBUS1:I2C:SEARCH:SETUP:NONACK:
READACCESS ON
:SERIALBUS1:I2C:SEARCH:SETUP:NONACK:
READACCESS?
-> :SERIALBUS1:I2C:SEARCH:SETUP:NONA
CK:READACCESS 1
:SERialbus<x>:I2C:SEARch[:SETup]:NON
ack:STARtbyte
Function Sets or queries whether or not NON ACK will be
ignored in start bytes for I
2
C bus signal searching.
Syntax
:SERialbus<x>:I2C:SEARch[:SETup]:NON
ack:STARtbyte {<Boolean>}
:SERialbus<x>:I2C:SEARch[:SETup]:NON
ack:STARtbyte?
<x> = 1 to 4
Example
:SERIALBUS1:I2C:SEARCH:SETUP:NONACK:
STARTBYTE ON
:SERIALBUS1:I2C:SEARCH:SETUP:NONACK:
STARTBYTE?
-> :SERIALBUS1:I2C:SEARCH:SETUP:NONA
CK:STARTBYTE 1
:SERialbus<x>:LIN?
Function Queries all LIN bus signal analysis and search
settings.
Syntax
:SERialbus<x>:LIN?
<x> = 1 to 4
:SERialbus<x>:LIN:ANALyze?
Function Queries all LIN bus signal analysis settings.
Syntax
:SERialbus<x>:LIN:ANALyze?
<x> = 1 to 4
:SERialbus<x>:LIN[:ANALyze]:SETup?
Function Queries all LIN bus signal analysis bus settings.
Syntax
:SERialbus<x>:LIN[:ANALyze]:SETup?
<x> = 1 to 4
:SERialbus<x>:LIN[:ANALyze]:SETup:BR
ATe
Function Sets or queries the LIN bus signal analysis bit
rate (data transfer rate).
Syntax
:SERialbus<x>:LIN[:ANALyze]:SETup:BR
ATe {<NRf>|USER,<NRf>}
:SERialbus<x>:LIN[:ANALyze]:SETup:BR
ATe?
<x> = 1 to 4
<NRf> = 1200, 2400, 4800, 9600, 19200
USER <NRf> = See the DLM4000 Features
Guide for this information.
Example
:SERIALBUS1:LIN:ANALYZE:SETUP:BRA
TE 2400
:SERIALBUS1:LIN:ANALYZE:SETUP:BRATE?
-> :SERIALBUS1:LIN:ANALYZE:SETUP:BRA
TE 2400
5.25 SERialbus Group

Table of Contents

Other manuals for YOKOGAWA DLM4038