5-201
IM DLM4038-17EN
Commands
5
:SERialbus<x>:PSI5[:ANALyze]:SETup:D
ATA:SNRejection
Function Queries all PSI5 signal analysis noise rejection
settings.
Syntax
:SERialbus<x>:PSI5[:ANALyze]:SETup:D
ATA:SNRejection?
<x> = 1 to 4
:SERialbus<x>:PSI5[:ANALyze]:SETup:D
ATA:SNRejection:ETIMe
Function Sets or queries the rejection end time for PSI5
signal analysis noise rejection.
Syntax
:SERialbus<x>:PSI5[:ANALyze]:SETup:D
ATA:SNRejection:ETIMe
:SERialbus<x>:PSI5[:ANALyze]:SETup:D
ATA:SNRejection:ETIMe?
<x> = 1 to 4
Example
:SERIALBUS1:PSI5:ANALYZE:SETUP:DATA:
SNREJECTION:ETIME 66.0000E-06
:SERIALBUS1:PSI5:ANALYZE:SETUP:DATA:
SNREJECTION:ETIME?
-> :SERIALBUS1:PSI5:ANALYZE:SETUP:DA
TA:SNREJECTION:ETIME 66.0000E-6
:SERialbus<x>:PSI5[:ANALyze]:SETup:D
ATA:SNRejection:MODE
Function Sets or queries the on/off status of the PSI5
signal analysis noise rejection.
Syntax
:SERialbus<x>:PSI5[:ANALyze]:SETup:D
ATA:SNRejection:MODE
:SERialbus<x>:PSI5[:ANALyze]:SETup:D
ATA:SNRejection:MODE?
<x> = 1 to 4
Example
:SERIALBUS1:PSI5:ANALYZE:SETUP:DATA:
SNREJECTION:MODE 1
:SERIALBUS1:PSI5:ANALYZE:SETUP:DATA:
SNREJECTION:MODE?
-> :SERIALBUS1:PSI5:ANALYZE:SETUP:DA
TA:SNREJECTION:MODE 1
:SERialbus<x>:PSI5[:ANALyze]:SETup:D
ATA:SOURce
Function Sets or queries the data source for PSI5 signal
analysis.
Syntax
:SERialbus<x>:PSI5[:ANALyze]:SETup:D
ATA:SOURce {<NRf>|MATH<y>}
:SERialbus<x>:PSI5[:ANALyze]:SETup:D
ATA:SOURce?
<x> = 1 to 4
<y> = 1 to 4
<NRf> = 1 to 8
Example
:SERIALBUS1:PSI5:ANALYZE:SETUP:DATA:
SOURCE 1
:SERIALBUS1:PSI5:ANALYZE:SETUP:DATA:
SOURCE?
-> :SERIALBUS1:PSI5:ANALYZE:SETUP:DA
TA:SOURCE 1
:SERialbus<x>:PSI5[:ANALyze]:SETup:N
UMBerofslot
Function Sets or queries the number of slots for PSI5
signal analysis.
Syntax
:SERialbus<x>:PSI5[:ANALyze]:SETup:N
UMBerofslot {AUTO|<NRf>}
:SERialbus<x>:PSI5[:ANALyze]:SETup:N
UMBerofslot?
<x> = 1 to 4
<NRf> = 1 to 6
Example
:SERIALBUS1:PSI5:ANALYZE:SETUP:NUMBE
ROFSLOT AUTO
:SERIALBUS1:PSI5:ANALYZE:SETUP:NUMBE
ROFSLOT?
-> :SERIALBUS1:PSI5:ANALYZE:SETUP:NU
MBEROFSLOT AUTO
:SERialbus<x>:PSI5[:ANALyze]:SETup:S
LOT<y>
Function Sets or queries the slot’s start and end times for
PSI5 signal analysis.
Syntax
:SERialbus<x>:PSI5[:ANALyze]:SETup:S
LOT<y> {<NRf>[,<NRf>]}
:SERialbus<x>:PSI5[:ANALyze]:SETup:S
LOT<y>?
<x> = 1 to 4
<y> = 1 to 6 (slot number)
<NRf> = 0 to 0.02 s (in 0.1 us steps)
Example
:SERIALBUS1:PSI5:ANALYZE:SETUP:SL
OT6 870us,1088us
:SERIALBUS1:PSI5:ANALYZE:SETUP:SL
OT1?
-> :SERIALBUS1:PSI5:ANALYZE:SETUP:SL
OT1870.000E-06,1.08800E−03
Description The first <NRF> is the slot's start time and the
second the end time.
If
:SERialbus<x>:PSI5[:ANALyze]:SETup
:NUMBerofslot
is a number from 1 to 6, when
the last slot number is specified with <y>, the
second <NRf> becomes valid.
If
:SERialbus<x>:PSI5[:ANALyze]:SETup:
NUMBerofslot
is
AUTO
, query is not possible.
:SERialbus<x>:PSI5[:ANALyze]:SETup:S
YNC
Function Sets or queries the sync signal for PSI5 signal
analysis.
Syntax
:SERialbus<x>:PSI5[:ANALyze]:SETup:S
YNC {<NRf>|MATH<y>|NONE}
:SERialbus<x>:PSI5[:ANALyze]:SETup:S
YNC?
<x> = 1 to 4
<y> = 1 to 4
<NRf> = 1 to 8
Example
:SERIALBUS1:PSI5:ANALYZE:SETUP:SY
NC 1
:SERIALBUS1:PSI5:ANALYZE:SETUP:SYNC?
-> :SERIALBUS1:PSI5:ANALYZE:SETUP:SY
NC 1
5.25 SERialbus Group