EasyManua.ls Logo

YOKOGAWA DLM4038 - Page 241

YOKOGAWA DLM4038
373 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-208
IM DLM4038-17EN
:SERialbus<x>:SENT[:ANALyze]:SETup:F
AST:USETup:DATA<y>:ORDer
Function Sets or queries the byte order of the specified
user-defined data of the SENT signal analysis
fast channel.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:
FAST:USETup:DATA<y>:ORDer {BIG|LITTl
e}
:SERialbus<x>:SENT[:ANALyze]:SETup:
FAST:USETup:DATA<y>:ORDer?
<x> = 1 to 4
<y> = 1 to 4
Example
:SERIALBUS1:SENT:ANALYZE:SETUP:FAST:
USETUP:DATA1:ORDER BIG
:SERIALBUS1:SENT:ANALYZE:SETUP:FAST:
USETUP:DATA1:ORDER?
->:SERIALBUS1:SENT:ANALYZE:SETUP:FAS
T:USETUP:DATA1:ORDER BIG
:SERialbus<x>:SENT[:ANALyze]:SETup:F
AST:USETup:DATA<y>:SIZE
Function Sets or queries the data size of the specified
user-defined data of the SENT signal analysis
fast channel.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:F
AST:USETup:DATA<y>:SIZE {<NRf>}
:SERialbus<x>:SENT[:ANALyze]:SETup:F
AST:USETup:DATA<y>:SIZE?
<x> = 1 to 4
<y> = 1 to 4
<NRf> = 0 to 24
Example
:SERIALBUS1:SENT:ANALYZE:SETUP:FAST:
USETUP:DATA1:SIZE 24
:SERIALBUS1:SENT:ANALYZE:SETUP:FAST:
USETUP:DATA1:SIZE?
-> :SERIALBUS1:SENT:ANALYZE:SETUP:FA
ST:USETUP:DATA1:SIZE 24
:SERialbus<x>:SENT[:ANALyze]:SETup:F
AST:USETup:MULTiplexing
Function Sets or queries the enable/disable condition of
the multiplexing for the user-defined data of the
the SENT signal analysis fast channel.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:
FAST:USETup:MULTiplexing {<Boolean>}
:SERialbus<x>:SENT[:ANALyze]:FAST:
USETup:MULTiplexing?
<x> = 1 to 4
Example
:SERIALBUS1:SENT:ANALYZE:SETUP:FAST:
USETUP:MULTIPLEXING ON
:SERIALBUS1:SENT:ANALYZE:SETUP:FAST:
USETUP:MULTIPLEXING?
-> :SERIALBUS1:SENT:ANALYZE:SETUP:
FAST:USETUP:MULTIPLEXING 1
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat?
Function Queries all SENT signal analysis format settings.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat?
<x> = 1 to 4
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:CEFactor?
Function Queries all SENT signal analysis error factor
settings.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:CEFactor?
<x> = 1 to 4
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:CEFactor:SAComm?
Function Queries all status and communication error factor
settings of SENT signal analysis.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:SAComm?
<x> = 1 to 4
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:CEFactor:SAComm:BIT<y>?
Function Queries all status and communication error factor
bit settings of SENT signal analysis.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:SAComm:BIT<y>?
<x> = 1 to 4
<y> = 0, 1
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:CEFactor:SCPulses
Function Sets or queries the consecutive calibration pulse
error factor of SENT signal analysis.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:
FORMat:CEFactor:SCPulses {OFF|OPT2|P
OPTion}
:SERialbus<x>:SENT[:ANALyze]:SETup:
FORMat:CEFactor:SCPulses?
<x> = 1 to 4
Example
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMA
T:CEFACTOR:SCPULSES OFF
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMA
T:CEFACTOR:SCPULSES?
-> :SERIALBUS1:SENT:ANALYZE:SETUP:FO
RMAT:CEFACTOR:SCPULSES OFF
5.25 SERialbus Group

Table of Contents

Other manuals for YOKOGAWA DLM4038