13
TI 11B08A01-01E
Ex db pxb II B + H
2
T4 Gb (programmed-temperature oven 95°C max., isothermal oven
95°Cmax.,liquid-samplevalve95°Cmax.)
*: Thesymbol“X”usedtodenotespecicconditionsofuse
Thesymbol“X”isusedtoprovideameansofidentifyingthatessentialinformationfortheinstallation,
use,andmaintenanceoftheequipmentiscontainedwithinthecerticate.
- Warning in the label refers you to read the instruction manual for an avoidance of static risk.
- Thegapbetweenameproofjointsisdifferentfromthestandardvaluesoftheameproof
standard.
Thejointdimensionsoftheameproofenclosuredifferfromtheminimumormaximumvalues
asa mentioned in the Ex d standard.
Sincethemodicationisnotallowed,thismanualgivesoutnodetailedinformation.
- All externally powered input signals into the pressurized enclosure shall be isolated by external
relays controlled by the Ex px protection system(Ex px safety device).
Any special conditions of use including particulars of possible misuse are shown below.
- Thethreadedtypeandsizeoftheameproofenclosureisonlyuses“M25x1.5”.
Thatisthereasonwhynoindicationontheameproofenclosure.
On the other hand, the threaded adapter has the indication of its own type and size.
- The protective gas shall be instrumental air.
- Temperature of protective gas (instrumental air) at the inlet of pressurized enclosure shall be
comply with the following temperature class range.
T1: -10 to +40°C
T2: -10 to +40°C
T3: -10 to +45°C
T4: -10 to +50°C
- OnlypersonnelauthorizedbyYokogawaElectricCorporationcanrepairtheequipment.
<GC8000-E (IECEx)>
IECExcertied:IECExDEK11.0083X*
Ex db pxb II B +H
2
T1 Gb (programmed-temperature oven 320°C max., isothermal oven
225°Cmax.,liquid-samplevalve250°Cmax.)
Ex db pxb II B + H
2
T2 Gb (programmed-temperature oven 225°C max., isothermal oven
225°Cmax.,liquid-samplevalve225°Cmax.)
Ex db pxb II B + H
2
T3 Gb (programmed-temperature oven 145°C max., isothermal oven
145°Cmax.,liquid-samplevalve145°Cmax.)
Ex db pxb II B + H
2
T4 Gb (programmed-temperature oven 95°C max., isothermal oven
95°Cmax.,liquid-samplevalve95°Cmax.)
*: Thesymbol“X”usedtodenotespecicconditionsofuse
Thesymbol“X”isusedtoprovideameansofidentifyingthatessentialinformationfortheinstallation,
use,andmaintenanceoftheequipmentiscontainedwithinthecerticate.
- Warning in the label refers you to read the instruction manual for an avoidance of static risk.
- Thegapbetweenameproofjointsisdifferentfromthestandardvaluesoftheameproof
standard.
Thejointdimensionsoftheameproofenclosuredifferfromtheminimumormaximumvalues
asa mentioned in the Ex d standard.
Sincethemodicationisnotallowed,thismanualgivesoutnodetailedinformation.
- All externally powered input signals into the pressurized enclosure shall be isolated by external
relays controlled by the Ex px protection system(Ex px safety device).
Any special conditions of use including particulars of possible misuse are shown below.
- Thethreadedtypeandsizeoftheameproofenclosureisonlyuses“M25x1.5”.
Thatisthereasonwhynoindicationontheameproofenclosure.
On the other hand, the threaded adapter has the indication of its own type and size.
- The protective gas shall be instrumental air.
- Temperature of protective gas (instrumental air) at the inlet of pressurized enclosure shall be
comply with the following temperature class range.
T1: -10 to +40°C
T2: -10 to +40°C
T3: -10 to +45°C
T4: -10 to +50°C
Sep. 07, 2017-00