1. Overview
40
TI 11B08A01-01E
1.3 Auxiliary Hardware
(1) Cylinders (carrier gas, standard gases, etc.)
Thesecylindersarelledwithacarriergasorstandardgas.
Themaximumllingpressureislimitedto15MPaconsideringthestrengthofthecylinders.
(2) Regulator for cylinder
This valve reduces a cylinder pressure to a safe degree to facilitate handling. The valve is directly
mounted to the cylinder.
(3) Stop valve
This valve shuts down the lines for the sample gas, instrument gas, and so on. It is operated
manually.
(4) Dehumidier
Any moisture in the carrier gas affects the columns. Therefore, if the carrier gas contains
moisture over 10 ppm, it is recommended to use a desiccant, such as a molecular sieve, to
prevent deterioration of the columns.
(5) Vent stacks
Thesearepipestodischargesamplebypassvent,samplevent,backushvent,foreushvent,
detector vent, measurement gas vent, and others to the atmosphere collectively.
Directtheexhaustgastoanareawherethegassufcientlydispersesanddiffusesbefore
discharging.
(6) Regulator for sample
Thisvalvereducesthesamplepressuretoaspecieddegree.
Tovaporizealiquidsample,usearegulatorwithasteam-heatedvaporizer.
(7) Pipings
The pipes for the sample inlet, carrier gas inlet, standard gas inlet, air for valve driving, air for
FID/FPD, FID/FPD hydrogen, steam, sampling bypass, and various vents are provided.
(8) Joints
The joints are used to connect pipes.
1.4 Gases Required for Operation
Thefollowinggasesarerequiredfortheanalyzer:
(1) Sample gas
The gas to be analyzed from the process line
Sep. 07, 2017-00