1. Overview
42
TI 11B08A01-01E
(6) FID/FPD combustion air
This air is used to burn hydrogen gas in an FID or FPD.
The air must satisfy the following conditions.
Purity: Measuring range from 0 to 50 ppm or more: moisture: 10 ppm or less; organic
components: 5 ppm or less
Measuring range from 0 to less than 50 ppm: moisture: 5 ppm or less; organic
components: 0.1 ppm or less
(7) Steam
Steamisrequiredtosteam-heatasample.Prepareasteamsourcethatcanapplythepressure
listed in “Operation Data.”
1.5 Conformance Standards
Safety Standard, EMC Standard and RoHS Standard are showned in the following list.
ExCertication MS code Safety Standard EMC standard RoHS standard
TIIS GC8000-T - - -
ATEX*1 GC8000-A EN 61010-1
EN 61010-2-030
GB 30439.1 (for NEPSI)
EN 61326-1 Class A, Table 2
EN 61326-2-3
Korea Electromagnetic
Conformity Standard
RCM Mark
EN 50581
IECEx*2 GC8000-E -
NEPSI*3 GC8000-P
FM GC8000-F
GC8000-G
FM 3810: 2005
(ANSI/ISA 61010-1-2004
(82.02.01),
ANSI/ISA-82.02.02-1996
(IEC 61010-2-031))
- -
CSA GC8000-C
GC8000-D
CSA C22.2 No. 61010-1-04 - -
*1: Analyzerbasesamplingunit(GCSMP)orbaseplacedunderGC8000-AisnotcoveredbytheEU-DoCnor
theECType-examinationCerticate.Theemptycompartment(base)placedbyYokogawatocreateaSelf-
standing GC8000-A, does not impair the compliance of the GC8000-A.
*2: ThedesignisbasedonthesafetyandEMCstandard,thoughthemarkofCEisnotindicated.
*3: ThedesignisbasedonthesafetyandEMCstandard,thoughthemarkofCE,KCandRCMarenot
indicated.
CAUTION
This instrument is a Class A product, and is designed for use in an industrial environment. Please
use this instrument in an industrial environment only.
Alistbelowshowstheapplicablestandardsforelectricalequipmentforexplosiveatmospheres.
Thislistdoesnotincludestandardsotherthanrelatedtoelectricalequipmentforexplosive
atmospheres.
Formoredetails,seeeachcerticate.PleasecontactyournearestYokogawarepresentative.
ExCertication MS code Standards for electrical equipment for explosive atmospheres
TIIS GC8000-T Constructionalrequirementsforelectricalequipmentforexplosive
atmospheres
Referenceguide;Typeexaminationguideforelectricalequipmentforexplosive
atmospheres (relating to technical standards conforming to international
standards) issued by the TIIS (Technical Institution of Industrial Safety), Japan.
(Issued in November, 1996.)
ATEX GC8000-A EN 60079-0:20012+A11, EN 60079-1:2014, EN 60079-2:2014
IECEx GC8000-E IEC 60079-0:2011, IEC 60079-1:2014, IEC 60079-2:2014
NEPSI GC8000-P GB 3936.1-2010, GB 3836.2-2010, GB 3836.5-2004
Sep. 07, 2017-00