5-5
IM MT300-02EN
Troubleshooting, Maintenance, and Inspection
5
Communication Query Errors (400 to 499)
Code Message Corrective Action Page
410 Query INTERRUPTED. Check the transmission and reception order. 10-2
*
420 Query UNTERMINATED. Check the transmission and reception order. 10-2
*
430 Query DEADLOCKED. Keep program messages to 1024 bytes or less in length,
including <PMT>.
10-2
*
440 Query UNTERMINATED after indefinite response. Do not write a query after *IDN? or *OPT?. —
* See the user’s manual (IM MT300-01EN), which is included in the accompanying CD.
Communication Warnings (50)
Code Message Corrective Action Page
50 *OPC/? exists in message. Write *OPC or *OPC? at the end of program messages. —
Other Errors (350)
Code Message Corrective Action Page
350 Queue overflow. Read the error queue. 12-6
*
* See the user’s manual (IM MT300-01EN), which is included in the accompanying CD.
Note
Code 350 occurs when the error queue overflows.
Execution Errors (800 to 899)
Code Message Corrective Action Page
800 Illegal date-time. Set the values correctly. 3-3
802 Invalid command. The command cannot be executed in user adjustment
mode.
—
* See the user’s manual (IM MT300-01EN), which is included in the accompanying CD.
System Errors (900/999)
Code Message Corrective Action Page
911 to
999
CPU board error. For details about the message and corrective action,
see section 5.2.
5-2
5.3 Communication Error Messages