1-185
IM 04L51B01-01EN
Configuring the GX/GP
1
1.21 Loading Settings
Load setting parameters and report templates, and load and delete scale images.
1.21.1 Loading Setting Parameters
Path
GX/GP: MENU key > Browse tab > Save load > Menu Load settings > Setting
parameters
Web application: Data tab > Save/Load > Setting parameters
Description
Setup Item Selectable Range or Options Default Value
Media kind SD, USB
1
Select file — —
1 Default value of Media kind
Media status Default Value
Non Media is not recognized
SD memory card SD
SD memory card, USB flash memory (/UH
option)
SD
USB meory (/UH option) USB
Media kind
Set the type of media.
Select file
Set the setting parameter file to load.
Setup Item Selectable Range or Options Default Value
All settings Off, On On
All settings
Set this to On to load all settings. Set this to Off to load security, IP address, and other
settings separately.
Setup Item Selectable Range or Options Default Value
Security
1
Off, On On
IP address
1
Off, On On
Other
1
Off, On On
Communication(Server settings)
2
Off, On On
Calibration correction settings
2
Off, On On
Instruments tags, etc.
2
Off, On On
1 Appears when All settings is set to Off.
2 Appears when Other is set to On. Release number 3 and later.
Security
Set this to On to load only the security settings.
Note
If the advanced security function (/AS option) is in use, passwords are not loaded except for
monitor users. All administrator and user passwords are set to their defaults.
See “Reusing Setting Parameters” section 1.3.3, “User Levels,” in the Advanced Security
Function (/AS) User’s Manual (IM 04L51B01-05EN).