1-207
IM 04L51B01-01EN
Configuring the GX/GP
1
8
Tap OK.
The settings are initialized. The message “Execution is complete.” will appear.
Operation complete
Note
• Before initializing, back up necessary setup parameters and other data.
Refer to page 1-196 in section 1.22.1, “Saving the Setting Parameters”
Explanation
Channel Assignment in Group Settings through Initialization
If you initialize, channels will be automatically assigned on the basis of the installed
modules. If necessary, change the channel assignments for your purpose.
1.25.2 Reconfiguring the GX/GP
When you reconfigure the GX/GP, the I/O modules installed in the GX/GP and expandable
I/O are detected, and the settings are changed accordingly.
Reconfiguration is necessary in the following situations.
• When you install a newly purchased module
• When you change the modules (change to different modules)
• When you add or remove modules
• When you connect or add an expandable I/O
If you purchased a I/O module preinstalled option (/U□□0 or /CR□□0), there is no need to
reconfigure if you are going to use the GX/GP as-is.
However, if you change modules, add modules, or delete modules, you will need to
reconfigure.
Path
GX/GP: MENU key > Browse tab > Initialize Calibration > Menu Reconfiguration
Procedure
1
Tap Execute.
The system information appears.
Switches the display
XXXXXXXXX
XX-XX-XX-XX-XX-XX
RX.XX.XX
RX.XX.XX
2
Tap Restructure.
A conrmation screen is displayed.
3
Tap OK.
The GX/GP will recongure.
Operation complete
1.25InitializingandCalibratingtheSystem(Initialization,reconfiguration,touchscreencalibration)