SHB_U14 Chapter 7: Configuration 7-43
Flow detector, continued
Start/span
correction –
Measurement range
correction
parameters
Position
Display
Offset correction
variable
Not applicable to auxiliary detectors
Ampl. variable Not applicable to auxiliary detectors
Offset fixed Start point correction, determined during an initial
calibration.
This value is only overwritten by a new initial calibration.
Ampl. fixed Sensitivity, determined during an initial calibration.
This value is only overwritten by a new initial calibration.
Start point desired
value
Air pressure display for initial zero calibration
Desired Span Value Air pressure display for initial span calibration
Start point raw value Raw value display (initial zero calibration)
Span raw value Raw value display (initial span calibration)
F
The flow detectors zero and span points must be calibrated.
Continued on next page