8-10 Chapter 8: Calibration
SHB_U14
Initial calibration of Uras 14 IR detectors, continued
Test
• The pressure detector should be calibrated
• Test all detectors (the measured values should be stable and regularly
updated)
• Status message 0 for all detectors
Performing the initial
calibration
Initial calibration is performed separately for each component. The large
measurement range should always be calibrated initially, the small range is
electronically determined from the large measurement range.
TCT calibrate (Enter the password)
Each component can be calibrated individually.
Right-clicking the measurement range opens the "Initial calibration" window.
Enter the desired values (actual test gas container values) and start the initial
calibration.
Continued on next page