13
Panel mounting (optional)
Referring to Figure 13:
1 Cut the correct sized hole in panel A.
2 Insert the transmitter into the panel cut-out B.
3 Screw one panel clamp anchor screw C into the left-hand
bracket D until 10 to 15 mm (0.39 to 0.59 in) of the thread
protrudes from the other side of the bracket and position
one clamp E over the end of the thread.
NOTICE
The correct torque is critical to ensure proper compression
of the panel seal and achieve the IPX6/NEMA 4X hosedown
6.
Dimensions in mm (in)
4 Holding assembly F together, position bracket D into the
left-hand recess on the rear of the transmitter and secure
with bracket securing screw G. Ensure that the plastic
washer remains in the position fitted.
5 Repeat steps 3 and 4 for the right-hand panel clamp
assembly.
6 Torque each panel clamp anchor screw to 0.5 to 0.6 Nm
(4.42 to 5.31 lbf/in).
Figure 13 Panel mounting the transmitter
A
B
C
D
E
F
A
G

max. thickness
6 (0.24)
Gasket
Panel cut-out dimensions
30 (1.2)
73 (2.87)
25.5
(1.00)
138
+1.0

+0.04

(5.43 )
138
+1.0

+0.04

(5.43 )
30
(1.2)