14 AWT210 | 2-WIRE CONDUCTIVITY, PH/ORP, PION TRANSMITTER | OI/AWT210-EN REV. C
…5 Mechanical installation
Transmitter installation
Pipe mounting (optional)
Referring to Figure 14, secure the transmitter to a pipe as
follows:
1 Fit two M6 x 50 mm hexagon-head screws A through one
clamp plate as shown.
2 Using the appropriate holes to suit vertical or horizontal
pipe, secure the clamp plate to the pipe-mounting bracket
B using two M6 x 8 mm hexagon-head screws and spring
lock washers C.
3 Position the pipe mounting bracket into the recesses on the
rear of the transmitter as shown and secure with the two
bracket securing screws D . Ensure the plastic washers
remain in the positions fitted.
4 Secure the transmitter to the pipe using the remaining clamp
plate, spring lock washers and nuts E.
Dimensions in mm (in)
NOTICE
If the optional weathershield F is used, locate it against
the transmitter back panel and attach the pipe-mount kit to
the weathershield rear face and transmitter.
A
C
D
E
C
B
B
E

Vertical
pipe
Horizontal
pipe
Pipe diameters:
max. 62 (2.44)/min. 45 (1.77)
Weathershield
(see page 12 for dimensions)
If weathershield is used, attach
mounting bracket to rear face
Figure 14 Pipe mounting the transmitter