34
AWT210 | 2-WIRE CONDUCTIVITY, PH/ORP, PION TRANSMITTER | OI/AWT210-EN REV. C
…11 Configuration
… Calibrate
Toroidal conductivity
Used to calibrate the sensor.
Note. Calibrate
transmitters connected to toroidal conductivity sensor types.
Access to the Calibrate menu is permitted via the Calibrate and Advanced levels or directly from
an Operator page using the Cal button.
Menu Comment Default
PV Zero Cal
PV Span Cal
Temperature Cal
Hold Output HART versions only. Not available if HART in Multidrop or Fixed Output mode.
Disabled/Enabled.
When Enabled: output is held automatically at the beginning of a PV calibration and released automatically at the end.
When Disabled: output remains active throughout any PV calibration.
Edit Cal
PV Slope
PV Offset
Temp Slope
Temp Offset
Restore Defaults
Calibra te
Exit
Menu
Select