AWT210 | 2-WIRE CONDUCTIVITY, PH/ORP, PION TRANSMITTER | OI/AWT210-EN REV. C
…11 Configuration
… Sensor Setup
2-electrode conductivity
Used to access standard setup parameters.
Note. Sensor Setup 
transmitters connected to 2-electrode conductivity sensor types.
Menu
Comment Default
Measurement Type Selects a measurement type compatible with the connected sensor:
Conductivity/Concentration
Cell Constant
Conc. Unit Sets the process value units.
Enabled only if Measurement Type = Concentration.
%/ppm/ppb/µg/l/mg/l
Conductivity Unit Enabled only if Measurement Type = Conductivity.
Auto/µS/cm/mS/cm
Conc. Curve Name Enabled only if Measurement Type = Concentration.
Conc. Curve Table Enabled only if Measurement Type = Concentration.
X1 Cond/Y1 Conc/X2 Cond/Y2 Conc/X3 Cond/Y3 Conc/
X4 Cond/Y4 Conc/X5 Cond/Y5 Conc/X6 Cond/Y6 Conc/Table Error
None
Not Saved
Temp. Unit Temperature unit displayed and applied to all temperature parameters/values.
°F/°C
Temp. Comp. Type Manual/Auto
Manual Temp. Enabled only if Temp. Comp. Type = Manual.
Auto TC Option Enabled only if Temp. Comp. Type = Auto.
Standard KCl/TC Coeff./Pure H2O/User Def. TC
TC Coeff. Enabled only if Temp. Comp. Type = Auto and Auto TC Option = TC Coeff.
 Enabled only if Temp. Comp. Type = Auto and Auto TC Option = Pure H2O.
Neutral/Acid./Base
User Def. TC Curve Enabled only if Temp. Comp. Type = Auto and Auto TC Option = User Def. TC.
X1 Temp/Y1 K/Kref/X2 Temp/Y2 K/Kref/X3 Temp/Y3 K/Kref/
X4 Temp/Y4 K/Kref/X5 Temp/Y5 K/Kref/X6 Temp/Y6 K/Kref/Table Error
None
Not Saved
Reference Temp.
Temp. Sensor Type Read-only.
3K Balco 2-wire/3K Balco 3-wire/PT 100 2-wire/PT 100 3-wire/PT 1000 2-wire/PT 1000 3-wire/None
Detect TC Type
TC Detection Status Read Only/Not Recognized/Recognized
Sensor Setup
Exit
Menu
Select