AWT210 | 2-WIRE CONDUCTIVITY, PH/ORP, PION TRANSMITTER | OI/AWT210-EN REV. C
15
6 Electrical installation
DANGER
If the transmitter is used in a manner not specified by the
Company, the protection provided by the equipment may
be impaired.
Refer to page 7 for electrical installation
considerations in Hazardous areas.
The transmitter conforms to Installation Category II of IEC
61010.
All equipment connected to the transmitter’s terminals
must comply with local safety standards (IEC 60950,
EN61010-1).
DANGER – CONNECTION/CABLE
REQUIREMENTS
The connection terminals accept cables with peripheral
wire cross-section of:
min.: 0.14 mm
2
(26 AWG)
max.: 1.5 mm
2
(14 AWG)
Do not use a rigid conductor material as this can result in
wire breaks.
Ensure the connecting cable is flexible.
To ensure the sensor cable length is sufficient, allow an
additional 100 mm (4 in) of cable to pass through cable
glands into the housing.
Ensure the correct connections are made to suit the
transmitter variant.
Terminal connections
Figure 15 Connections overview
123
4
5
123
4
5
AB AB
++
123
4
5
6
7
8
Power Ammeter
SENSE
GUARD
RTD 1
RTD 2
SHIELD
RTD 3
REF
SOL_GND
12345678
12345678
12345678
DRIVE +
DRIVE –
RTD 1
RTD 2
SHIELD
RTD 3
DRIVE +
DRIVE –
RTD 1
RTD 2
SHIELD
RTD 3
SENSE +
SENSE –
DRIVE +
DRIVE –
RTD 1
RTD 2
SHIELD
RTD 3
SENSE +
SENSE –
Communication module connections
Sensor module connections
HART modules
HART modules
2-electrode conductivity modules
4-electrode conductivity modules
Toroidal conductivity modules
pH/ORP/plon modules