AWT210 | 2-WIRE CONDUCTIVITY, PH/ORP, PION TRANSMITTER | OI/AWT210-EN REV. C
21
8 Diagnostic alarms
Conguration
Failure
Check Function
Off Specication
Maintenance Required
Alarm code in the format:
--XPPP.NNN--
X = Alarm class
F = Failure
C = Check function
S = Off specification
M = Maintenance required
PPP = Alarm priority
(high number = high priority)
NNN = Alarm number
Alarm class
Alarm message
(uppercase)
Alarm group
Recovery action
(lowercase)
Check temperature
sensor and perform
TC recognition
--S010.047--
MANUAL TEMPERATURE
COMPENSATION MODE
Back Exit
Figure 19 Example diagnostic alarm
Note.
Alarms are listed in alarm priority order (high number = high priority alarm).
Diagnostic message ALARM MESSAGE Recovery action pH

conductivity

conductivity
Toroidal
conductivity
HART FF PA
Electronics

SENSOR MODULE
MEMORY FAILURE
Change sensor module
Electronics

COMMS MODULE
MEMORY FAILURE
Change comms module
Electronics

CURRENT OUTPUT
NOT CALIBRATED
Trim output
If problem persists
change comms module
Conguration

DATA
SIMULATION
Conguration

CURRENT OUTPUT
FIXED
Enable loop current
mode. Disable loop test/trim
& PV cal.
Process

CURRENT OUTPUT
SATURATED
Adjust engineering
range
Electronics

SENSOR MODULE
FAILURE
Change sensor module
Process

OPEN CABLE OR SENSOR
OUT OF SOLUTION
Check sensor wiring
Verify that sensor
is in solution
Electronics

PRIMARY VARIABLE
INPUT READ ERROR
Check sensor
If problem persists
change sensor module
Electronics

2ND PRIMARY VARIABLE
INPUT READ ERROR
Check sensor
If problem persists
change sensor module
Operation

SHORTED CABLE OR
GROUND LOOPS PRESENT
Check sensor wiring
Sensor

SENSOR
POLARIZATION
Check process
Check sensor wiring
Clean sensor
Table 2 Diagnostic alarms