…8 Diagnostic alarms
Diagnostic message ALARM MESSAGE Recovery action pH

conductivity

conductivity
Toroidal
conductivity
HART FF PA
Process

SENSOR IS DIRTY
Clean sensor
Electronics

DIAGNOSTIC
INPUT READ ERROR
Check terminals
Check sensor wiring
Check electrode
Electronics

LOW ELECTRODE
IMPEDANCE
Check terminals
Check sensor wiring
Check electrode
Process

PRIMARY VARIABLE
OUTSIDE PHYS. LIMITS
Check sensor wiring
Check configuration
Process

PRIMARY VARIABLE
OUTSIDERANGE LIMITS
Adjust engineering
range
Electronics

TEMPERATURE
INPUT READ ERROR
Check sensor
If problem persists
change sensor module
Process

SENSOR TEMPERATURE
OUTSIDE LIMITS
Check sensor wiring
Check temperature
configuration
Sensor

HIGH SENSOR
EFFICIEN C Y (slo p e)
Check calibration
Clean sensor
Check sensor wiring
Sensor

LOW SENSOR
EFFICIEN C Y (slo p e)
Check calibration
Clean sensor
Check sensor wiring
Sensor

HIGH SENSOR
OFFSET
Check calibration
Clean sensor
Check sensor wiring
Sensor

LOW SENSOR
OFFSET
Check calibration
Clean sensor
Check sensor wiring
Electronics

DIAGNOSTIC
INPUT READ ERROR
Check sensor wiring
If problem persists
change sensor module
Electronics

REFERENCE IMPEDANCE
INPUT READ ERROR
Check sensor
If problem persists
change sensor module
Sensor

HIGH REFERENCE
IMPEDANCE
Check sensor
Check sensor wiring
Operation

POWER SUPPLY
VOLTAGE
OUTSIDE LIMITS
Trim output
Ensure power supply
voltage is within
limits
Electronics

SENSOR MODULE
VOLTAGE WARNING
Check sensor wiring
If problem persists
change sensor module
Conguration

MANUAL TEMPERATURE
COMPENSATION MODE
Check temperature
sensor and perform
TC recognition
Table 2 Diagnostic alarms