AWT210 |  2-WIRE CONDUCTIVITY, PH/ORP, PION TRANSMITTER  |  OI/AWT210-EN REV. C
…11 Configuration
… Easy Setup
4-electrode conductivity
Used to setup standard parameters quickly.
Note. Easy Setup 
transmitters connected to 4-electrode conductivity sensor types.
Menu Comment Default
Language  Selects the language to be displayed on screen:
English/Deutsch/Français/Español/Italiano
Note.   The language options are displayed with localized spelling.
Measurement Type Selects a measurement type compatible with the connected sensor:
Conductivity/Concentration
Sensor Group A/B
Conc. Solution Enabled only if Measurement Type = Concentration.
0-15% NaOH/0-20% NaCl/0-18% HCl/0-20% H2SO4/User Defined       
Conc. Unit Sets the process value units.
Enabled only if Measurement Type = Concentration.
%/ppm/ppb/µg/l/mg/l
Conc. Curve Name Enabled only if Measurement Type = Concentration.
Temp. Unit Temperature unit displayed and applied to all temperature parameters/values.
Not enabled if Temperature Compensation Type = Manual.  
°F/°C 
 Selects range of values displayed on Operator Page 1 for examples of character formats/sizes.
PV/PV + Temperature/PV + Temp + PV%/PV + Temp + O/P
Easy Setup
Exit
Menu