28 AWT210 | 2-WIRE CONDUCTIVITY, PH/ORP, PION TRANSMITTER | OI/AWT210-EN REV. C
11 Configuration
Easy Setup
pH
Used to setup standard parameters quickly.
Note. Easy Setup
transmitters connected to pH sensor types.
Menu Comment Default
Language Selects the language to be displayed on screen:
English/Deutsch/Français/Español/Italiano
Note. The language options are displayed with localized spelling.
Measurement Type Selects a measurement type compatible with the connected sensor:
pH/ORP/Ion Conc/pION
pH Sensor Type Select a measurement type compatible with the connected sensor:
Glass/Antimony/Custom
Isopotential Pt. Isopotential point.
Enabled only if Measurement Type = pH and pH Sensor Type = Custom.
Asymmetric Pot. Asymmetric potential.
Enabled only if Measurement Type = pH and pH Sensor Type = Custom.
PV Unit Sets the process value units.
Enabled only if Measurement Type = Ion Conc.
%/ppm/ppb/µg/l/mg/l
Valence The ion valence determines the millivolt change per decade of concentration.
Enabled only if Measurement Type = Ion Conc.
–3/–2/–1/1/2/3
Magnitude
Enabled only if Measurement Type = Ion Conc. 1/2/3
End Magnitude The ion concentration state functions by associating an end millivolt
value to an end magnitude value.
Enabled only if Measurement Type = Ion Conc.
10/100/1000
End mV The end millivolt value associated with the end magnitude value.
Enabled only if Measurement Type = Ion Conc.
Temp. Unit Temperature unit displayed and applied to all temperature parameters/values.
Not enabled if Temperature Compensation Type = Manual.
°F/°C
Temp. Comp. Type .
Manual/Auto/Auto Solution
Note. When Auto is selected and a temperature sensor is recognized, the temperature compensation
calculation is based on the Nernst equation detailed on . When Auto Solution is selected and a
temperature sensor is provided with the probe, the calculation is based on the Nernst solution coefficient
equation detailed in detailed on . When Manual is enabled/selected, the value set in the Manual
Temp. parameter is used for the referenced temperature compensation.
Auto
(if valid
temperature
sensor
recognized)
Manual Temp. Temperature value used for compensation in Temp Comp. Type/Manual compensation mode.
Enabled only if Temperature Compensation Type = Manual.
pH Soln. Coeff. Enabled only if Measurement Type = pH and Temp. Comp. Type = Auto Solution.
mV Soln. Coeff. Enabled only if Measurement Type = pH and Temp. Comp. Type = Auto.
Selects range of values displayed on Operator Page 1 for examples of character formats/sizes.
PV/PV + Temperature/PV + Temp + PV%/PV + Temp + O/P
Easy Setup
Exit
Menu
Select