AWT210 | 2-WIRE CONDUCTIVITY, PH/ORP, PION TRANSMITTER | OI/AWT210-EN REV. C
35
Sensor Setup
pH
Used to access standard setup parameters.
Note. Sensor Setup
transmitters connected to pH sensor types.
Menu Comment Default
Measurement Type Selects a measurement type compatible with the connected sensor:
pH/ORP/Ion Conc/pION
pH Sensor Type Enabled only if Measurement Type = pH.
Select a measurement type compatible with the connected sensor:
Glass/Antimony/Custom
PV Unit Enabled only if Measurement Type = Ion Conc.
Sets the process value units.
%/ppm/ppb/µg/l/mg/l
Isopotential Pt. Isopotential point.
Enabled only if Measurement Type = pH and pH Sensor Type = Custom.
Asymmetric Pot. Asymmetric potential.
Enabled only if Measurement Type = pH and pH Sensor Type = Custom.
Valence The ion valence determines the millivolt change per decade of concentration.
Enabled only if Measurement Type = Ion Conc.
–3/–2/–1/1/2/3
Magnitude
concentration output.
Enabled only if Measurement Type = Ion Conc.
1/2/3
End Magnitude The ion concentration state functions by associating an end millivolt value to an end magnitude value.
Enabled only if Measurement Type = Ion Conc.
10/100/1000
End mV The end millivolt value associated with the end magnitude value.
Enabled only if Measurement Type = Ion Conc.
Temp. Unit Temperature unit displayed and applied to all temperature parameters/values.
Not enabled if Temperature Compensation Type = Manual.
°F/°C
Temp. Comp. Type .
Manual/Auto/Auto Solution
Note. When Auto is selected and a temperature sensor is recognized, the temperature compensation calculation
is based on the Nernst equation detailed on . When Auto Solution is selected and a temperature
sensor is provided with the probe, the calculation is based on the Nernst solution coefficient equation detailed
in detailed on . When Manual is enabled/selected, the value set in the Manual Temp. parameter is used
for the referenced temperature compensation.
Auto
(if valid
temperature
sensor
recognized)
Manual Temp. Temperature value used for compensation when Temp Comp. Type = Manual compensation mode.
Enabled only if Temperature Compensation Type = Manual.
pH Soln. Coeff. Enabled only if Measurement Type = pH and Temp. Comp. Type = Auto Solution.
mV Soln. Coeff. Enabled only if Measurement Type = pH and Temp. Comp. Type = Auto.
Temp. Sensor Type Read-only.
3K Balco 2-wire/3K Balco 3-wire/PT 100 2-wire/PT 100 3-wire/PT 1000 2-wire/PT 1000 3-wire/None
Detect TC Type
TC Detection Status Read Only/Not Recognized/Recognized
Sensor Setup
Exit
Menu
Select