AWT210 |  2-WIRE CONDUCTIVITY, PH/ORP, PION TRANSMITTER  |  OI/AWT210-EN REV. C
43
1-point manual calibration
Performs a manual calibration (Offset adjustment) at a single 
reference point.
1 Wait for stable reading
–                          – – – 
PV         7.00 pH
Continue When Stable
 
1 Point Manual
Abort
Continue
Monitor the process value and 
continue ( ) to the next step once 
the value has stabilized.
2 Enter the new value
PV         7.00 pH
    New        
0  0 7.1 2  
1 Point Manual
Next
Continue
Enter the desired PV value by 
pressing the   key to move the 
cursor and the     keys to 
change the value. When the new 
value has been entered press the 
 key to continue.
3 Completion
Following a successful calibration 
the calculated coefficients are 
displayed.
Following an unsuccessful 
calibration the reason for failure 
is displayed.
2-point manual calibration
Performs a 2-point calibration using 2 pre-defined buffer 
solutions.
1 Buffer temperature 
–                          – – – 
Buffer Temperature
25.0 °C 
2 Point Manual
Next
Edit
The temperature of the buffer 
solutions is displayed. The 
temperature can be edited by 
pressing the   key. When the 
buffer temperature is correct press 
the   key to continue.
2 Buffer 1 value
–                          – – – 
Buffer 1
4.00 pH 
2 Point Manual
Next
Edit
The value of the 1
st
 buffer solution 
is displayed. The value can be 
edited by pressing the   key.  
When the buffer value is correct 
press the   key to continue.
3
Wait for stable reading – 1
st
 
buffer solution
–                          – – – 
PV          4.03 pH 
2 Point Manual
Abort
Continue
Immerse In Buffer 1  
Continue When Stable
Immerse the sensor in the buffer 
solution, monitor the process value 
and continue   to the next step 
once the value has stabilized.
4 Buffer 2 value
–                          – – – 
Buffer 2
7.00 pH 
2 Point Manual
Next
Edit
The value of the 2
nd
 buffer solution 
is displayed. The value can be 
edited by pressing the   key.  
When the buffer value is correct 
press the   key to continue.
5
Wait for stable reading –  
2
nd
 buffer solution
–                          – – – 
PV         7.15 pH 
2 Point Manual
Abort
Continue
Immerse In Buffer 2  
Continue When Stable
Immerse the sensor in the buffer 
solution, monitor the process value 
and continue   to the next step 
once the value has stabilized.
6 Completion
Following a successful calibration 
the calculated coefficients are 
displayed.
Following an unsuccessful 
calibration the reason for failure 
is displayed.
–                          – – – 
1 Point Manual
Exit
PV       7.12 pH
PV Slope     100.00% 
PV Offset      8.94 mV
1 Point Manual
Exit
PV    7.14   pH 
Calibration Failed 
Offset Too High
–                          – – – 
2 Point Manual
Exit
PV        7.00 pH
PV Slope      98.75% 
PV Offset      8.94 mV
2 Point Manual
Exit
PV        7.15 pH 
Calibration Failed 
Slope Too High