EasyManua.ls Logo

ABB AWT210 - Conductivity Sensor Module Connections; 2-Electrode Sensors; 4-Electrode Sensors; Toroidal Sensors

ABB AWT210
54 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AWT210 | 2-WIRE CONDUCTIVITY, PH/ORP, PION TRANSMITTER | OI/AWT210-EN REV. C
17
Conductivity sensor module connections
2-electrode sensors
Sensor type RTD wiring
DRIVE +
DRIVE –


SHIELD




 Red Black
Green/
Yellow Blue
Brown
 Red Black Brown
Green/
Yellow
Blue
  Green Black Blue Yellow Dark green

 Green Black Blue/Red Yellow Dark green
 Green Black Yellow Red Dark green Blue
NOTICE
AWT210 2-electrode conductivity sensor modules are supplied
standardized to theoretical sensor characteristics. Following
installation, but before use, a process calibration should be
performed to ensure optimum accuracy. For 2-electrode
conductivity sensor calibration procedures see page 44.
4-electrode sensors
Sensor type RTD wiring
DRIVE +
SENSE +
SENSE –
DRIVE –


SHIELD

  Green Red White Black Blue Yellow Dark green
NOTICE
AWT210 4-electrode conductivity sensor modules are supplied
standardized to theoretical sensor characteristics. Following
installation, but before use, a process calibration should be
performed to ensure optimum accuracy. For 4-electrode
conductivity sensor calibration procedures see page 45.
Toroidal sensors
Sensor type RTD wiring
DRIVE +
DRIVE –
SENSE +
SENSE –


SHIELD

  Black Blue White Red Green Yellow Dark green
NOTICE
AWT210 toroidal conductivity sensor modules are supplied
standardized to theoretical sensor characteristics. Following
installation, but before use, a process calibration should be
performed to ensure optimum accuracy. For toroidal
conductivity sensor calibration procedures see page 45.