EasyManua.ls Logo

ABB FSM4000-SE41F - Page 45

ABB FSM4000-SE41F
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrical connections
D184B140U02 FSM4000 45
Pulse output
Pulse output active Pulse output, passive, optocoupler
R
B
*
52
G00247
51
R
B
*
U
CE
I
CE
+ 16 ... 30 V DC
-
I E
Function:
fmax:
fmin:
Setting range:
active/passive, selectable via jumper (see "Commissioning" in operating instructions)
5 kHz,
0.00016 Hz
Pulses/unit, pulse width (observe dynamic limits)
Active:
150 Load < 10 k Pulse width 50 ms, max. pulse frequency 3 Hz,
500 Load < 10 k Pulse width 0.1 ms, max. pulse frequency: 5 kHz
Passive:
"closed": 0 V UC
EL
2 V, 2 mA I
CEL
220 mA
"open": 16 V U
CEH
30 V, 0 mA I
CEH
2 mA
Terminals: 51, 52
Fig. 43: Pulse output, active and passive, optocoupler
PROFIBUS PA / FOUNDATION Fieldbus
G00248
2
FF-
FF+
R
C
1
97
PA-
R
C
PA+
98
97
98
I EI E
The resistance R and capacitor C form the bus termination. They
must be installed when the device is connected to the end of the
entire bus cable.
R = 100 ; C = 1 µF
1PROFIBUS PA
2FOUNDATION Fieldbus
I = Internal
E = external
Fig. 44