Parameterization
D184B140U02 FSM4000 85
Submenu/parameter Value range Comment
As Startup Copies the current data record in the Fingerprint "as
Startup" (commissioning fingerprint).
Accept Yes, exit via C/CE
Yes → Enter
Submenu
Calibration
Submenu Enter the current coil temperature to readjust the coil
resistance at 20 °C. Perform adjustment only on a "cold"
device. See also chapter "7.4 Additional information
regarding use of enhanced diagnostic functions".
Coil Temp
Temp. in °C -100 ... 100
0.0
Submenu Enter the length of the signal cable.
Note: Information must be entered to calculate the coil
resistance / coil temperature.
CabLength
Length [m] 0 ... 200
0.0
Submenu Offset correction for temperature display.
Temp. Offset
Temp. in °C -100 ... 100
0.0
Import RT20 Import the reference resistance for the coil.
See also chapter "7.4 Additional information regarding
use of enhanced diagnostic functions".
Note: This menu is displayed only when the coil
resistance or coil temperature are enabled as an error or
warning message.
Accept Yes, exit C/CE
New 500
ohms
Old 450 ohms
Yes → Enter
Electr. Zero Point This adjustment is required in order to use the diagnostic
measurement values for the electrode voltage E1/E2 and
the electrode balance as well as the wiring errors for the
electrode state.
Electr. Zero
Point
Yes, exit C/CE Display the current (new) and stored (old) zero values for
electrode E1/E2.
Select "Enter" to import the current zero points
Note: The fluid in the flowmeter sensor must be at
absolute standstill.
New 150 180 µV
Old 200 190
µV
Yes → Enter
Electr. Analysis Displays stored electrode zero and current electrode AC
voltage acc. to amount and phase.
Where
N1: Current measured electrode AC voltage E1
N2: Current measured electrode AC voltage E2
A1: Stored zero E1
A2: Stored zero E2
N1: 12 µV 286° Yes, exit C/CE
N2: 10 µV 283°
A1: 650 µV 133°
A2: 670 µV 135°