EasyManua.ls Logo

abi 7500 - Page 54

abi 7500
260 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 Design the Relative Standard Curve Experiment
Set Up the Standards
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
30
Notes
Design
Guidelines
When you design your own relative standard curve experiment:
Set up a standard curve for each target in the reaction plate. The targets are
previously defined in the Targets screen (“Set Up the Targets” on page 26).
Enter the number of points for each standard curve in the reaction plate. Applied
Biosystems recommends at least five dilution points for each standard curve.
Enter the number of identical reactions (replicates) for each point in the standard
curve. Applied Biosystems recommends three replicates for each point.
Because the range of standard quantities affects the amplification efficiency
calculations, carefully consider the appropriate range of standard quantities for your
assay:
For more accurate measurements of amplification efficiency, use a broad range of
standard quantities, such as between 10
5
and 10
6
. If you specify a broad range of
quantities for the standards, you need to use a PCR product or a highly
concentrated template, such as a cDNA clone.
If you have a limited amount of cDNA template and/or if the target is a low-copy
number transcript, or known to occur within a specified range, a narrow range of
standard quantities may be necessary.
The serial factor is used to calculate the quantities in all points of the standard curve.
If your starting quantity is the highest quantity, select a serial factor such as 1:2, 1:3,
and so on. If your starting quantity is the lowest quantity, select a concentration
factor such as 2, 3, and so on.
1
2
5
If needed, use the scroll bar to view GAPDH,
then perform steps 4a and 4b.
3b
3a

Table of Contents