Symbols, problems and error messages
You have inserted a used test strip into the meter.
Start again from the beginning with an unused test strip.
The measuring window is dirty.
Clean the measuring window (see Chapter 8).
You did not insert the test strip correctly or completely.
Hold the test strip so that the arrows printed on it and the orange square face up-
wards. Without bending it, gently push the test strip into the test strip guide in the
direction of the arrows until you feel it lock into place.
You applied blood or control solution to the test strip too soon, i.e. before the drop
symbol ashed in the display.
Repeat the test with a new test strip.
The test strip was bent or moved while testing was in progress.
Repeat the test with a new test strip.
The meter was exposed to a strong electromagnetic eld.
Change your position or turn off the source of radiation.

Table of Contents

Other manuals for Accu-Chek ACTIVE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek ACTIVE and is the answer not in the manual?

Accu-Chek ACTIVE Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1 - 2 μL
Test Time5 seconds
Memory Capacity500 results
Display TypeLCD
Automatic CodingYes
Blood SampleCapillary whole blood
Dimensions97.8 x 46.8 x 19.1 mm
Battery Type1 x 3V lithium battery (type CR2032)
Battery LifeApproximately 1000 tests
Data TransferVia infrared (IR) to computer
Operating Conditions8-42°C, 10-85% RH