Battery Critical Alarm
Medium Priority Alarm - This alarm screen appears about 30 minutes before the battery power runs out.
It is indicated by a repeating audible tone.
Press Audio Pause to silence the audible tone.
The audible tone will come back on in two
minutes if the alarm condition has not been
fixed.
Therapy will continue during this alarm.
However, if the unit is not plugged in
within 30 minutes, therapy will turn off.
This alarm is fixed by recharging the battery.
1. Connect the therapy unit to a wall outlet using the
power supply. Refer to the Battery Charging
Instructions section for more information.
2. Press Reset to return to the Patient Mode
Home
screen.
3. Make sure therapy is on. The On / Off
button will be green.
4. Press On / Off to restart therapy if needed.
WARNING: If the therapy unit will be off
for more than two hours, call your doctor
or nurse right away. Without power to the
therapy unit, your dressing will need to be
replaced.
Battery critically low; charge immediately
Battery Critical
Audio Pause
Reset
Battery Low