74
Specifications
Specifications subject to change without notice.
Dimensions: .......................................................................................................................... 7.6 in W x 6 in H x 2.5 in D
(19.3 x 15.2 x 6.4 cm)
Weight (with empty canister attached): ....................................................................................~2.4 lbs (~1.08 kg)
Pressure Options: ................................................................................................................................. 25 to 200 mmHg
(3.3 to 26.6 kPa)
Therapy Delivery Modes: .............................................................................................. Continuous or Intermittent
Canister Volume: ...................................................................................................................................................~300 mL
Electrical:
Battery Run Life:............................................................................................. ~14 hours, depending on settings
Battery Charge Time: ............................................................................~6 hours from a fully discharged state
External Power Supply Input: .....................................................................................................100-240VAC 0.8A
50 - 60 Hz
External Power Supply Output: .................................................................................................................12V, 3.3 A
Patient & Enclosure Leakage Current: ......................................................................................<100 Microamps
Environmental Conditions (60601 2nd Edition):
Storage / Transport Conditions
Temperature Range: ...............................................................................................-4°F (-20°C) to 140°F (60°C)
Relative Humidity Range: .......................................................................................... 0-95% non-condensing
Operating Conditions
Temperature Range: ..................................................................................................41°F (5°C) to 104°F (40°C)
Relative Humidity Range: .......................................................................................15 - 95% non-condensing
Atmospheric Pressure: ...........................................................................................................1060 hpa to 700 hpa
Environmental Conditions (60601 3rd Edition):
Storage / Transport Conditions
Temperature Range: .............................................................................................-13°F (-25°C) to 158°F (70°C)
Relative Humidity Range: .......................................................................................... 0-93% non-condensing
Operating Conditions
Temperature Range: ..................................................................................................41°F (5°C) to 104°F (40°C)
Relative Humidity Range: .......................................................................................15 - 93% non-condensing
Atmospheric Pressure: ...........................................................................................................1060 hpa to 700 hpa
Expected Service Life ......................................................................................................................................... 3 years
IEC Classification (60601 2nd Edition)
Medical Equipment
Equipment not suitable for use in presence of flammable anesthetic mixture with air, oxygen or
nitrous oxide.
Type B, Applied Part
Class I
IPX0
IEC Classification (60601 3rd Edition)
Medical Equipment
Equipment not suitable for use in presence of flammable anesthetic mixture with air, oxygen or
nitrous oxide.
Type BF, Applied Part
Class II
IP22 - Protection against solid objects greater than 12.5 mm and against liquid water falling for
short periods of time.
The dressing components of the V.A.C.® Therapy System are considered Applied Parts under IEC 60601-1 Third Edition.