30
Therapy Inactive Alarm
Medium Priority Alarm - This alarm screen appears when therapy has been off for 15 minutes (with the
unit powered on). This alarm will be accompanied by a repeating audible tone.
Press Audio Pause to silence the audible tone.
The audible tone will come back on in two
minutes if the alarm condition has not been
fixed.
This alarm is fixed by restarting therapy.
1. Press Reset to return to the Patient Mode
Home
screen.
2. Press On / Off to restart therapy.
If therapy is not desired, turn the therapy unit
off using the Power button on the front of
the unit.
WARNING: If the therapy unit will be off
for more than two hours, call your doctor
or nurse right away. Without power to the
therapy unit, your dressing will need to be
replaced.