PATIENT
27
Blockage Alarm Therapy Interrupted
Medium Priority Alarm - This alarm screen appears when the tubing is blocked. This alarm will be
accompanied by a repeating audible tone.
Press Audio Pause to silence the audible tone.
The audible tone will come back on in two
minutes if the alarm condition has not been
fixed.
Therapy will continue during this alarm.
However, therapy may not be at the proper
pressure.
This alarm is fixed by checking the tubing.
1. Make sure both clamps on the tubing are open.
2. Make sure that the tubing is not kinked, crimped or
blocked in any way.
3. If the Blockage Alarm Therapy Interrupted remains after
completing steps 1 and 2, lower the therapy unit and
tubing to level with or below the wound site.
4. Press Reset to return to the Patient Mode
Home
screen.
5. If the alarm persists or you cannot find the blockage,
call KCI or your doctor or nurse.
WARNING: If the therapy unit will be off
for more than two hours, call your doctor
or nurse right away. Without power to the
therapy unit, your dressing will need to be
replaced.
WARNING:
Pump may be running, but therapy
is interrupted. Press ‘?’ for more information.
Audio Pause
Reset
Blockage
Blockage Alarm
Therapy Interrupted