25
5. Make sure therapy is on. The On / Off button will be
green.
6. Press On / Off to restart therapy if
needed.
7. If the leak alarm persists or you cannot find the leak, call
KCI or your doctor or nurse.
Use the SEAL CHECK™ Leak Detector feature:
You can also find a leak using the SEAL CHECK™ feature
(see SEAL CHECK
TM
Leak Detector section). Press SEAL
CHECK™ on the
Leak Alarm
screen to use the SEAL CHECK™
Leak Detector.
Leak Alarm Therapy Interrupted
Medium Priority Alarm - This alarm screen appears when the Leak Alarm has not been fixed and
therapy has stopped. It is indicated by a repeating audible tone.
Press Audio Pause to silence the audible
tone. The audible tone will come back on in
two minutes if the alarm condition has not
been fixed.
This alarm is fixed by finding and fixing the leak.
1. Fix any leaks as described in the Leak Alarm section.
2. Press Reset to return to the Patient Mode
Home
screen.
3. Press On / Off to restart therapy.
4. If the alarm persists or you cannot find the leak, call KCI
or your doctor or nurse.
WARNING: If the therapy unit will be off for more than two hours,
call your doctor or nurse right away. Without power to the therapy
unit, your dressing will need to be replaced.
If the leak condition is not fixed, the Leak Alarm will reappear.
Continue troubleshooting the leak as described in the previous
section.