Appendix I User Access Quick Reference
92
OsmoTECH XT Single-Sample Micro-Osmometer User Guide
System Functions User Access Levels
Administrator Supervisor Operator
• MAINTENANCE
– Clean the sample probe and chamber
9
9
9
– Perform and record preventative maintenance
9
8
8
– Perform and record instrument repairs
9
8
8
• Set the installation date
9
8
8
RESULTS screen
• Send results to LIS
• PRINT results
• Display STATISTICS for up to 20 results
• Display a control monitoring CHART for selected control
• Filter results
9 9 9
• EXPORT results to CSV or PDF
• Review and approve results
• Add comments to results records
9
9
9
SETTINGS screen
• DATE & TIME (including SNTP)
• LANGUAGE
9
9
9
• USERS
– Create user accounts
9
9
8
– Delete or disable user accounts (includes disabling the Service account)
9
9
8
• Edit user accounts
– Assign user access features
– Change password
– Unlock account
9
9
8
• Set up connection to an LDAP server
9
9
8
• HW OPTIONS
– Audio click tone and volume
– Screen brightness
– Enable/disable barcode scanner
9
9
9
• PROBE RESISTANCE (change block and sample probe resistance)
9
9
9
• CONTROLS
– Establishing control limits
– Editing control information
– Deleting controls
– Displaying a Control Monitoring chart
9
9
9
– Set up warnings for a control
9
9
8