Repair and Service
Safety Guidelines / General Repair Instruction
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
3 Safety Notes for Service Activities
3.1 General Safety Notes for all Service Activities
CAUTION:
Sharp edges inside the digitizer: Cut or abrasion possible.
Be careful at any maintenance work.
CAUTION:
Image plate or cassette unit can be damaged.
Do not insert and pull out cassette as long as the digitizer is switched off.
WARNING:
120 mW Laser (Class 3B) in the Optic Module may cause eye injuries.
Avoid direct and indirect eye contact.
Do not keep tools in the laser beam when the device is switched on.
Do not open the cover of the Optic Module.
3.2 Safety Notes for Installation Planning
Warning:
Images can be lost due to power failure.
Connect the equipment to an un-interruptible power supply (UPS) or an institutional
standby generator.
WARNING:
When different combinations of equipment are used in various medical
environments a potential difference (V) can exist between the protective earths
in different localities. If the protective earthing fails this potential difference can
cause a HAZARD for the OPERATOR or for the PATIENT.
To comply with ISO 60601-1 (annex I) all computers and peripherals must be
connected to the same power source.
Always connect the associated monitor to the same Uninterruptible Power
Source (UPS) as the PC.
Edition 1, Revision 2 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 3.1 / Page 6 of 11
03-2013 Type 5151 / 100 / 110 / 200 / 210 Agfa Company Confidential