Repair and Service
Troubleshooting
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
6.2 Dust Stripes
Symptom
Dust stripes are visible in the clinical
image or a flatfield image.
Figure 4: Example for dust stripes
Cause
Dirt or dust in the scan path.
Solution
(1) Use the cleaning brush to clean the mirror and the entrance surface of the light
collector by moving the brush a few times. For instructions refer to the digitizer
user manual.
(2) Expose and scan a flatfield and evaluate again.
For instructions how to create a flatfield image refer to chapter 3.6, Adjustments
and Calibrations
(3) If unacceptable image flaws suntil occur: Escalate to the next support level.