Functional Description
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
4 Boot-up of the Digitizer
The following flowchart illustrates the boot-up of the digitizer:
Digitizer idle
Error message
in Error Viewer.
Boot-up stops
Message
“Digitizer start-up”
(4198) in Error Viewer
Drawer Motor
briefly
audible
Error message
in Error Viewer.
Boot-up stops
Wait till the Digitizer
receives "ready for
identification” message
Boot-up of PMI and
Power Board finished
Load data and
check E-labels
Warning message
in Error Viewer
Rescue Cycle: Digitizer
tries to solve problem.
IP Clean-up
Cycle
Set drawer motor
to HOME
Selftest okay?
Initialization okay?
NX Viewer
running?
Cassette or IP
found?
NO
NO
NO
YES
YES
YES
NO
YES
Digitizer Default Boot-Up Sequence
pproximate
time in
seconds
Visible / a
Effects
udible
1 sec.
0
17
40
Status LED
green on
Status LED slow red blinking (approx. 0,6 sec. off / 0,6 sec. on)
515102bb.cdr
43
100*
105*
* time depends on
alvo ad
ustment time. Can be up to 5 minutes.
Adjust galvo
60
Digitizer
Switch-On
Status LED starts triple red blinking
if no network connection to NX.
fast red blinking
if no connection to Error Viewer.
Status LED starts
Send Self Test
Signal to all SW modules
Figure 10
Edition 1, Revision 1 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 2 / Page 12 of 19