DAP meter
Model Diamentor CI-P
Type Device for the dose area product meas-
urement in X-ray diagnostics according
to IEC 60580 standard.
Principle of measurement Radiation measure with ionization
chamber
Measured quantity Dose area product
Digital resolution 0.01 μGym
2
Max linearity error < 2.5%
Nominal range of dose-area prod-
uct rate
(0.01 - 2500) μGym
2
/s
Equivalent filtration of the cham-
ber @75kV
0.3 mm Al
Max measurement field 118 mm x 118 mm
Dimensions (W x D x H) 152 mm x 234 mm x 23 mm
Weight 455 g
DR 100e (Digital Radiography) | Product Information | 111
0365B EN 20181015 0853