On the NX application
1. Check if the DR Detector Switch displays the name of the DR Detector
that's being used
2. If a wrong DR Detector is displayed, select the right DR Detector by
clicking the drop down arrow on the DR Detector Switch.
Related Links
DR Detector Switch on page 22
On a DR Detector that has a status indicator
Check if the status of the DR Detector is ready for exposure. If the status is
not ready for exposure, the DR Detector cannot be used for making an
exposure.
On the X-ray generator control panel
Check if the exposure settings displayed on the control panel are suitable
for the exposure.
If other exposure values are required than those defined in the NX exam,
use the control panel to overwrite the default defined exposure settings.
Related Links
X-ray generator control panel on page 25
On the control panel
WARNING:
Check if the led beacon light lights up green.
This indicates that the wireless DR Detector and the X-ray generator are
ready for making an exposure.
Related Links
Led beacon light on page 30