Warning messages
Warning messages must be reset using the reset button in order to continue
working.
Figure 13: Reset button
Message Description Action
CLOCK ERR. System clock error Press the reset button to
continue.
INI. APR APR checksum error Press the reset button to
continue.
ERROR IN APR An APR value is out of
range
Change the value.
TUBE SEASONING After a long idle period
(3 months or more) it is
necessary to perform X-
ray tube seasoning, to
avoid damage.
Press the reset button to
continue. Call Service
for the tube seasoning.
HOT TUBE The temperature of the
X-ray tube has reached
the maximum allowed
value.
Wait for the monobloc
to cool down.
EXPIRED TIME The exposure button
has been pressed half-
way for preparing the X-
ray tube for more than
15 seconds.
Release the exposure
button and repeat the X-
ray exposure sequence.
MANUAL STOP The exposure button
has been released be-
fore the end of the ex-
posure.
Press the reset button to
continue.
INACTIVE DAP Dosimeter not connec-
ted.
-