accordance with IEC
61000-4-11
U
r
) for 25 peri-
ods at 0°
0% U
r
for 250 pe-
riods
If the user wants the device
to work continuously, even
when the energy supply is
interrupted, it is recommen-
ded to use an energy supply
free of interruptions or a
battery.
Magnetic field at the
supply frequency
(50/60 Hz) in accord-
ance with IEC
61000-4-8
30 A/m Magnetic field at the net-
work frequency should cor-
respond to the typical values
as they are in a
commercial and clinical en-
vironment.
REMARK : U
r
is the alternating current in the network before the application
of the test level.
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
Tests of Resistance to Dis-
ruption
Test level of profes-
sional medical
equipment and ba-
sic EMC standards
Electromagnetic Envi-
ronment
Recommended protec-
tive distance:
Conducted high frequency
disturbance variables in ac-
cordance with IEC
61000-4-6
6 V rms in ISM fre-
quency bands be-
tween 150 kHz and
80 MHz
Radiated high frequency
disturbance variables in ac-
cordance with IEC
61000-4-3
3 V/m
80 MHz to 2.7 GHz
RF communication 28 V/m
Disruptions are possi-
ble near devices that
carry the following
symbol:
The field strength of stationary transmitters, such as base stations of radio
telephones, mobile broadcasts for rural areas, amateur stations, and AM and
DR 100e (Digital Radiography) | Product Information |
119