132 Agilent 1200 Preparative Pump User Manual
7 Maintenance
Simple Repairs
Replacing the Module’s Firmware
To upgrade/downgrade the module’s firmware the following steps have to be
performed:
1 Download the required module firmware, the latest LAN/RS-232 FW
Update Tool and the documentation from the Agilent web.
http://www.chem.agilent.com/scripts/cag_firmware.asp.
2 Load the firmware into the module as described in the documentation.
When The installation of newer firmware might be necessary
if a newer version solves problems of older versions or
to keep all systems on the same (validated) revision.
The installation of older firmware might be necessary
to keep all systems on the same (validated) revision or
if a new module with newer firmware is added to a system or
if third part control software requires a special version.
Tools required LAN/RS-232 Firmware Update Tool or
Agilent Diagnostic Software
Instant Pilot G4208A (only if supported by module)
Control Module G1323B (only if supported by module)
Parts required Description
Firmware, tools and documentation from Agilent web site
Preparations Read update documentation provided with the Firmware Update Tool.
NOTE
Due to a different hardware platform, there is no way to convert a G1314D VWD and the
G1314E VWD SL Plus to a G1314A/B VWD or G1314C VWD SL.