Agilent 1200 Preparative Pump User Manual 97
Troubleshooting and Diagnostics
6
Pressure Test
Evaluating the Results
The sum of all leaks between the pump and the blank nut will be indicated by
flow rate of > 10 µl/minute to keep the pressure constant. Note that small
leaks may cause the test to fail, but solvent may not be seen leaking from a
module.
Large internal leaks at the EMPV can easily be identified by observing a flow
out through the waste tubing of the EMPV, while the EMPV should be closed.
NOTE
Please notice the difference between an error in the test and a failure of the test! An error
means that during the operation of the test there was an abnormal termination. If a test
failed, this means that the results of the test where not within the specified limits.
NOTE
Do not overtighten fittings! After the first major increase of resistance when tightening a
previously loose fitting, the fitting should not be tightened any further than an additional
1/8 to 1/4 wrench turn.
NOTE
Often it is only a damaged blank nut itself (poorly shaped from overtightening) that causes
a failure of the test. Before investigating on any other possible sources of failure make sure
that the blank nut you are using is in good condition and properly tightened!

Table of Contents

Other manuals for Agilent Technologies 1200 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 series and is the answer not in the manual?

Agilent Technologies 1200 series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals