Agilent 1200 Preparative Pump User Manual 97
Troubleshooting and Diagnostics
6
Pressure Test
Evaluating the Results
The sum of all leaks between the pump and the blank nut will be indicated by 
flow rate of > 10 µl/minute to keep the pressure constant. Note that small 
leaks may cause the test to fail, but solvent may not be seen leaking from a 
module.
Large internal leaks at the EMPV can easily be identified by observing a flow 
out through the waste tubing of the EMPV, while the EMPV should be closed.
Please notice the difference between an error in the test and a failure of the test! An error 
means that during the operation of the test there was an abnormal termination. If a test 
failed, this means that the results of the test where not within the specified limits.
Do not overtighten fittings! After the first major increase of resistance when tightening a 
previously loose fitting, the fitting should not be tightened any further than an additional 
1/8 to 1/4 wrench turn.
Often it is only a damaged blank nut itself (poorly shaped from overtightening) that causes 
a failure of the test. Before investigating on any other possible sources of failure make sure 
that the blank nut you are using is in good condition and properly tightened!